Background: Drug-coated balloons (DCBs), which deliver anti-proliferative drugs with the aid of excipients, have emerged as a new endovascular therapy for the treatment of peripheral arterial disease. In this study, we evaluated the use of keratose (KOS) as a novel DCB-coating excipient to deliver and retain paclitaxel. Methods: A custom coating method was developed to deposit KOS and paclitaxel on uncoated angioplasty balloons. The retention of the KOS-paclitaxel coating, in comparison to a commercially available DCB, was evaluated using a novel vascular-motion simulating ex vivo flow model at 1 h and 3 days. Additionally, the locoregional biological response of the KOS-paclitaxel coating was evaluated in a rabbit ilio-femoral injury model at 14 days. Results: The KOS coating exhibited greater retention of the paclitaxel at 3 days under pulsatile conditions with vascular motion as compared to the commercially available DCB (14.89 ± 4.12 ng/mg vs. 0.60 ± 0.26 ng/mg, p = 0.018). Histological analysis of the KOS–paclitaxel-treated arteries demonstrated a significant reduction in neointimal thickness as compared to the uncoated balloons, KOS-only balloon and paclitaxel-only balloon. Conclusions: The ability to enhance drug delivery and retention in targeted arterial segments can ultimately improve clinical peripheral endovascular outcomes.
The purpose of this pilot study was to investigate the feasibility of a perfusion catheter to deliver liquid paclitaxel into arterial segments. A clinically relevant rabbit ilio-femoral injury model was utilized to determine the impact of liquid paclitaxel delivered locally into the vessel wall using a perfusion catheter at 1 h to 14 days. Treatment by two clinically available forms of liquid paclitaxel, a solvent-based (sb) versus an albumin-bound (nab), along with a control (uncoated balloons), were investigated. Pharmacokinetic results demonstrated an increase in the retention of the sb-paclitaxel versus the nab-paclitaxel at 1 h; however, no other differences were observed at days one, three, and seven. Histological findings at 14 days showed significantly less neointimal area in the sb-paclitaxel treated arteries as compared with the nab-paclitaxel and the uncoated balloon-treated arteries. Additionally, percent area stenosis was significantly less in the sb-paclitaxel group. These results support the concept of local liquid delivery of paclitaxel into the arterial segments.