OBJECTIVE. Inappropriate patient centering on the gantry changes the size of the localizer radiographs used for CT examinations, influencing the operation of CT automatic tube current modulation because tube current is controlled with information from localizer radiographs. The purpose of this study was to examine the influence of inappropriate patient centering on the gantry isocenter on automatic tube current modulation.MATERIALS AND METHODS. An elliptical phantom was scanned with four automatic tube current modulation techniques after acquisition of localizer radiographs in the horizontal and vertical directions with the phantom center shifted from the gantry isocenter in the vertical direction. After scanning, the magnification rate of the frontal localizer radiographs, tube current–time product, and image noise were examined.RESULTS. On phantom studies, the magnification rate of localizer radiographs showed a linear relation to the vertical deviation of the phantom from the gantry isocenter. From 50 ...
Recently, specific computed tomography (CT) scanners have been equipped with organ‐based tube current modulation (TCM) technology. It is possible that organ‐based TCM will replace the conventional dose‐reduction technique of reducing the effective milliampere‐second. The aim of this study was to determine if organ‐based TCM could reduce radiation exposure to the breasts without compromising the image uniformity and beam hardening effect in thoracic CT examinations. Breast and skin radiation doses and the absorbed radiation dose distribution within a single section were measured with an anthropomorphic phantom and radiophotoluminescent glass dosimeters using four approaches to thoracic CT (reference, organ‐based TCM, copper shielding, and the combination of the above two techniques, hereafter referred to as the combination technique). The CT value and noise level were measured using the same calibration phantom. Organ‐based TCM and copper shielding reduced radiation doses to the breast by 23.7% and 21.8%, respectively. However, the CT value increased, especially in the anterior region, using copper shielding. In contrast, the CT value and noise level barely increased using organ‐based TCM. The combination technique reduced the radiation dose to the breast by 38.2%, but greatly increased the absorbed radiation dose from the central to the posterior regions. Moreover, the CT value increased in the anterior region and the noise level increased by more than 10% in the entire region. Therefore, organ‐based TCM can reduce radiation doses to breasts with only small increases in noise levels, making it preferable for specific groups of patients, such as children and young women.PACS numbers: 87.53.Bn; 87.57.Q‐; 87.57.qp
6 January 2014 in Journal of Applied Clinical Medical Physics
Determination of the half-value layer (HVL) and quarter-value layer (QVL) values is not an easy task in X-ray computed tomography (CT), because a nonrotating X-ray tube must be used, which requires the assistance of service engineers. Therefore, in this study, we determined the accuracy of the lead-covered case method, which uses X-rays from a rotating X-ray tube, for measuring the HVL and QVL in CT. The lead-covered case was manufactured from polystyrene foam and a 4 mm thick lead plate. The ionizing chamber was placed in the center of the case and aluminum filters were placed 15 cm above the aperture surface. Aperture widths of 1.0, 2.0, and 3.0 cm for a tube voltage of 110 kV and an aperture width of 2.0 cm for the tube voltages of 80 and 130 kV were used to measure exposure doses. The results of the HVL and QVL were compared with those of the conventional nonrotating method. A 2.0 cm aperture was believed to be adequate, because of its small differences in the HVL and QVL in the nonrotating method and its reasonable exposure dose level. When the 2.0 cm aperture was used, the lead-covered case method demonstrated slightly larger HVLs and QVLs (0.03-0.06 mm for the HVL and 0.2-0.4 mm for the QVL) at all the tube voltage settings. However, the differences in the effective energy were 0.1-0.3 keV; therefore, it could be negligible in an organ-absorbed dose evaluation and a quality assurance test for CT. PACS numbers: 87.57.-s; 87.57.Q-; 87.57.uq