22 April 2015 in Micromachines
Verified
Journal Article
Open access
PDF file
TPTom Platenkamp
Tom Platenkamp
WSWouter Sparreboom
Wouter Sparreboom
GRGijs Ratering
Gijs Ratering
MKMarcel R. Katerberg
Marcel R. Katerberg
JLJoost Lötters
Joost Lötters
This article describes a primary calibration setup, and its uncertainty, for low flow liquid calibrations at Bronkhorst High-Tech. It will be used to calibrate reference flow meters from 1 to 200 g/h. By setting up an uncertainty budget for this setup, the calibration of the instruments can be compared to that of NMI’s (National Metrology Institutes). The uncertainty budget consists of mass, time and mass flow uncertainties/corrections that need to be taken in to account for determining the traceable mass flow. Tests results of different flow meters/actuators measured on the setup support the calculated uncertainty. By participating in an intercomparison with NMI’s the measurement and uncertainty of this setup is traceable to European NMI’s.