This study presents high-precision analyses of calcium (Ca) isotope ratios using a collision-cell multi-collector inductively coupled plasma mass spectrometer (CC-MC-ICP-MS, Nu Sapphire). This instrument allows Ca isotope measurement in the low resolution mode, requiring much less sample compared with TIMS and conventional MC-ICP-MS, and has additional measurement capability of both radiogenic (ε40Ca) and stable isotope ratios (δ44/40Ca and δ44/42Ca). The effects of several factors on the precision and accuracy of Ca isotopic data were evaluated, including the total Ca concentration, Ca intensity mismatch between the sample and standard, HNO3 molarity mismatch between the sample and standard, and the presence of matrix elements. High precision was obtained on 100 ng g−1 Ca samples (i.e., for 6 replicates ∼600 ng of Ca). Accurate measurements require the following conditions: (1) Ca intensities matched within 3% between the sample and standard, (2) low contents of Sr, Mg and Cr (Sr/Ca, Mg/Ca and Cr/Ca ratios need to be lower than 0.01%, and K/Ca, Al/Ca, Ti/Ca and Cr/Ca ratios <1%). Eleven standard reference materials analyzed using the established method yield highly consistent Ca isotope results with literature values. The long-term external precisions are higher than 0.07‰ (2 SD) for δ44/40Ca and higher than 0.05‰ (2 SD) for δ44/42Ca. The slope of the fractionation line between δ44/40Ca and δ44/42Ca is 1.9616 ± 0.0794, within the error of the slope value from the exponential law.
30 July 2022 in Journal of Analytical Atomic Spectrometry
Here we present an analytical method for Fe isotopic measurements in low resolution mode, using a Nu Sapphire collision cell-equipped multi-collector inductively coupled plasma mass spectrometer by standard-sample bracketing. High Fe sensitivity was obtained through reducing Ar based interference by over 9 orders of magnitude whilst other analytes remained unaffected. The effects of Fe concentration, Fe beam intensity and HNO3 molarity mismatch between the standard and sample and the presence of matrix elements have been evaluated. The long-term analyses of JMC Fe and BCR-2 indicate that the obtained isotopic ratios are highly reproducible, with precisions of better than ±0.03‰ for δ56Fe (2SD). Such accurate and precise data could be acquired via 3 to 6 repeat measurements consuming 50–100 ng Fe, which has improved by a factor 40 compared to those previously reported on other instruments. Accurate measurements were achieved by closely matching Fe intensities between the sample and standard (a 5% mismatch would create a 0.02‰ offset in the 56Fe/54Fe ratio). On the other hand, the ratio of the concentration between the matrix elements and Fe, such as CNa/CFe, CMg/CFe, CCa/CFe, CTi/CFe and CNi/CFe, should be kept under 0.1, CK/CFe under 0.3, CAl/CFe and CCu/CFe under 1, and CCo/CFe and CZn/CFe under 2 to avoid any matrix effect. In addition, high precision Fe isotopic data were obtained on twenty-one geological reference materials and were highly consistent with the literature values. Furthermore, we obtained δ56Fe of standard chalcopyrite (XTC) for the first time: 0.127 ± 0.027‰ (2SD, n = 3).
17 January 2023 in Journal of Analytical Atomic Spectrometry
With single-column chemical purification, highly accurate δ7 Li values for eight international reference materials are presented together with seventeen Chinese geological reference materials from NRCG using the Nu Sapphire.
23 January 2023 in Journal of Analytical Atomic Spectrometry
Here we present an analytical method for Cu isotopic measurement using standard-sample bracketing on a Nu Sapphire multi-collector inductively coupled plasma mass spectrometer. The effects of Cu concentration and HNO3 molarity mismatch between the standard and the sample, as well as the presence of matrix elements have been evaluated. Intermediate precision of better than ±0.030‰ for δ65Cu (2SD) was routinely obtained on NIST SRM 3114 and NWU-Cu-A. Accurate measurements were achieved when CNa/CCu < 0.5, CTi/CCu, CFe/CCu and CCo/CCu < 1, CNi/CCu < 3 and CZn/CCu < 10. High-precision Cu isotopic data were determined for nineteen geological reference materials, including igneous rocks, sedimentary rocks, metamorphic rocks and chalcopyrite, with an overall range of ∼0.44‰ (−0.056‰ to 0.384‰), nearly 15 times the current analytical precision. The Cu isotopic data of these reference materials can be used for future interlaboratory comparisons.