This paper describes the realization of a new system for resistance scaling from 10 ${\rm k}\Omega$ to 100 ${\rm T}\Omega$. The resistance unit has been transferred from the quantum hall resistance (QHR) primary resistance standard to the resistance standard with a nominal value of 10 ${\rm k}\Omega$ by a cryogenic current comparator. For further transfer of the resistance unit from 10 ${\rm k}\Omega$ to 100 ${\rm T}\Omega$, Hamon transfer devices (Hamon networks) are used. For its complete realization five devices have been developed: 10–100–1000 ${\rm k}\Omega$, 1–10–100 ${\rm M}\Omega$, 0.1–1–10 ${\rm G}\Omega$ , 10–100–1000 ${\rm G}\Omega$, and 1–10–100 ${\rm T}\Omega$. The first two devices have a single insulation, and the next three have a double isolation. In the transfer devices with the double insulation, triax-type connectors are used. All transfer devices have individual temperature stabilization within ${\pm}{\rm 0.01}^{\circ}{\rm C}$. In the paper, the factors influencing device accuracy have been described. The insulation leakage is recognized, and the error caused by it is calculated. The guarding network for the new design of Hamon transfer devices is presented. Voltage coefficients of resistance and settling times of resistors used in these devices have been determined and discussed. Uncertainties for the Hamon transfer devices ratios have been calculated. Calibration results have been presented. The developed system is currently tested in the Central Office of Measures in Poland.
This paper describes a resistance transfer system based on resistance transfer devices. Authors developed a single path resistance transfer system and tested it in the Central Office of Measures in Poland. Although the system was useful it had some disadvantages, such as the lack of temperature stabilization of secondary standards, voltage influence, resistance unit transfer from the primary standard to secondary standards only with one path. To eliminate or minimize those disadvantages, authors are working on the development of a double path resistance transfer system, which is described in this paper.
Hamon transfers allow accurate comparison of resistance value of standards in 1:10 and 1:100 ratios. In the paper idea of construction design and development of the Hamon transfers, used for resistance unit transfer from the primary standard QHR (Quantum Hall Resistance) to high value resistance standards up to 100 TΩ, in which factors influencing transfers uncertainty were minimized, has been shown.
In this paper, a guarded active-arm bridge for high-resistance standards with programmable two-channel source of DC voltage is described. Both channels of this source are controlled from a common reference voltage. This ensures a high stability of the voltage ratio of both source channels. The presented results of tests of this source show that its use in the active bridge is more advantageous than the use of the two independent DC voltage sources. The bridge allows high accuracy in the comparison of the high-resistance standards in 1:1 ratio. The current state of knowledge about active-arm bridges is presented, the basics of the bridge operation are given, the two-channel DC voltage source used in the bridge and its test results are described, the measurement process that does not require transposition of the compared resistance standards is described, sample measurement results, and a plan for further work are given. The bridge is used in the resistance standard calibration system at the Central Office of Measures (GUM), Poland, in the range from 100 MΩ to 100 TΩ.
This paper describes realization in Poland of new system for resistance unit transfer from 10 kΩ to 100 TΩ, based on Hamon transfers. The new design of those transfers and their properties has been presented. Currently this system is being tested in Central Office of Measures in Poland.