Asynchronous read-out single-photon avalanche diode (SPAD) array detectors1 – 3 are expected to terrifically extend the abilities of fluorescence laser-scanning microscopy (LSM). In contrast with the typical LSM single-element detectors, but similar to scientific cameras, this class of sensors preserves the spatial distribution of the impinging fluorescence photons. Differently from a scientific camera, every element of the array is a fully independent SPAD,4 which provides a high temporal precision () digital signal for each detected single photon. The combination of tailored data acquisition (DAQ) systems with these SPAD array detectors enables photon-resolved measurement of the fluorescence emission: every photon can be tagged with (i) spatial signatures, namely, the emission and detection coordinates of the photon in the detector and sample plane, respectively, and (ii) temporal signatures, namely, the time delay of the photon with respect to specific reference events. A series of recent publications5 – 10 have demonstrated the benefits of leveraging the unique information provided by the SPAD arrays. The photon-resolved spatial and temporal information improves the performance of most advanced LSM-based techniques and paves the way for new ones. Such unprecedented results laid the foundation of a new microscopy paradigm that we named single-photon laser-scanning microscopy (SP-LSM).9 The features of a particular technique based on the SP-LSM paradigm depend on the characteristics of the SPAD array detectors11,12 and those of the DAQ systems. The spatial information alone, provided by the detector array, has already boosted the performance of numerous techniques. Among them, confocal LSM evolved into image-scanning microscopy (ISM) capable of generating superresolved images at an exceptional signal-to-noise ratio (SNR) level.6,13 – 15 Furthermore, detector arrays also boosted the optical sectioning capabilities of ISM.10 Importantly, the same benefits have been easily transported to two-photon-excitation16 and stimulated-emission depletion (STED) microscopy.10 Fluorescence-fluctuation spectroscopy17 (FFS) is another technique that has benefited from structured detection,3,8 gaining access to molecular mobility information previously hidden. The aforementioned techniques require access to temporal scales of the order of the pixel dwell time or (histogram) bin time width for imaging or FFS, respectively. These experimental parameters are typically in the range of a few microseconds. Indeed, the relative DAQ systems only need to count photons—on every channel of the SPAD array—detected within each pixel of the image or bin of the histogram. However, the power of the SP-LSM paradigm is fully unleashed with the techniques requiring access to the nano- or picosecond scale. In fluorescence lifetime ISM (FLISM),6,9 the photon arrival time with respect to the fluorophore excitation event enables mapping the fluorescence lifetime distribution in the sample at superresolution. Such information helps us understand the properties of the biomolecular environment,18 decipher biomolecule structural changes,19 and implement multispecies imaging.20 In quantum ISM (Q-ISM),5 the photon coincidences allow the construction of an image with a spatial resolution beyond the limits of conventional ISM. Superresolution optical fluctuation ISM7 exploits fluorescence fluctuations on the (sub-)microsecond scale to improve spatial resolution. In these cases, the DAQ system must include a multichannel time-tagger (TT) module. This latter registers each photon detection event with spatial and temporal coordinates, namely, the position on the detector plane and the detection times—absolute and relative to a series of reference signals (e.g., the sync of the laser pulses and the sync from the beam scanning system). Time-tagger DAQ (TT-DAQ) systems are the greatest ally for SP-LSM, fully preserving the single-photon information collected by the SPAD array detectors. However, they are more complex and expensive than photon-counting DAQ systems. Furthermore, they require transferring and storing a large amount of data. TT-DAQ typically records signals with tens of picoseconds of precision, with a range of up to several seconds. Although the complexity and cost can be reduced using DAQ boards based on field-programmable gate array (FPGA),9,21 the storage problem cannot be solved without sacrificing some information. A possible solution would be to design an application-specific TT-DAQ system, where the temporal characteristics and information content are bounded by the needs of the measurements, as has recently been done for Q-ISM.22 In this work, we propose a valuable alternative to high-precision time-tagging DAQ systems to implement FLISM. We leverage the digital-frequency-domain (DFD) principle23,24 to implement an FPGA-based multichannel DAQ system able to reduce data transfer and storage load without compromising the performance of FLISM. Specifically, we use a heterodyne scheme to record the photon arrival time with a sampling period down to 400 ps, which grants excellent sampling of the decay curve of the most common fluorophores. Instead of transferring the arrival time of each detection event to the computer, we build the fluorescence decay histogram directly on the FPGA board. Thus we only need to transfer and store a histogram for each channel and imaging pixel (or histogram time bin for FFS). This strategy also enables real-time visualization of the lifetime measurements and embedding on the same FPGA board as the time-resolving DAQ and the microscope control system.