Purpose: To develop a time‐efficient pulse sequence that acquires multiple diffusion‐weighted images with distinct diffusion times in a single shot by using multiple stimulated echoes (mSTE) with variable flip angles (VFA).Methods: The proposed diffusion‐weighted mSTE with VFA (DW‐mSTE‐VFA) sequence begins with two 90° RF pulses that straddle a diffusion gradient lobe (GD) to excite and restore one half of the magnetization into the longitudinal axis. The restored longitudinal magnetization was successively re‐excited by a series of RF pulses with VFA, each followed by another GD, to generate a set of stimulated echoes. Each of the multiple stimulated echoes was acquired with an EPI echo train. As such, the train of multiple stimulated echoes produced a set of diffusion‐weighted images with varying diffusion times in a single shot. This technique was experimentally demonstrated on a diffusion phantom, a fruit, and healthy human brain and prostate at 3 T.Results: In the phantom experiment, the mean ADC measured at different diffusion times using DW‐mSTE‐VFA were highly consistent (r = 0.999) with those from a commercial spin‐echo diffusion‐weighted EPI sequence. In the fruit and brain experiments, DW‐mSTE‐VFA exhibited similar diffusion‐time dependence to a standard diffusion‐weighted stimulated echo sequence. The ADC showed significant time dependence in the human brain (p = 0.003 in both white matter and gray matter) and prostate tissues (p = 0.003 in both peripheral zone and central gland).Conclusion: DW‐mSTE‐VFA offers a time‐efficient tool for investigating the diffusion‐time dependency in diffusion MRI studies.
Purpose: To develop a DWI sequence with multiple readout echo‐trains in a single shot (multi‐readout DWI) over a reduced FOV, and to demonstrate its ability to achieve high data acquisition efficiency in the study of coupling between diffusion and relaxation in the human prostate.Methods: The proposed multi‐readout DWI sequence plays out multiple EPI readout echo‐trains after a Stejskal‐Tanner diffusion preparation module. Each EPI readout echo‐train corresponded to a distinct effective TE. To maintain a high spatial resolution with a relatively short echo‐train for each readout, a 2D RF pulse was used to limit the FOV. Experiments were performed on the prostate of six healthy subjects to acquire a set of images with three b values (0, 500, and 1000 s/mm2) and three TEs (63.0, 78.8, and 94.6 ms), producing three ADC maps at different TEs and three maps at different b values.Results: Multi‐readout DWI enabled a threefold acceleration without compromising the spatial resolution when compared with a conventional single‐readout sequence. Images with three b values and three TEs were obtained in 3 min 40 s with an adequate SNR (≥ 26.9). The ADC values (1.45 ± 0.13, 1.52 ± 0.14, and 1.58 ± 0.15 ; P < 0.01) exhibited an increasing trend as TEs increased (63.0 ms, 78.8 ms, and 94.6 ms), whereas values (74.78 ± 13.21, 63.21 ± 7.84, and 56.61 ± 5.05 ms; P < 0.01) decreases as the b values increased (0, 500, and 1000 s/mm2).Conclusion: The multi‐readout DWI sequence over a reduced FOV provides a time‐efficient technique to study the coupling between diffusion and relaxation times.
It has been increasingly reported that in biological tissues diffusion-weighted MRI signal attenuation deviates from mono-exponential decay, especially at high b-values. A number of diffusion models have been proposed to characterize this non-Gaussian diffusion behavior. One of these models is the continuous-time random-walk (CTRW) model, which introduces two new parameters: a fractional order time derivative α and a fractional order spatial derivative β. These new parameters have been linked to intravoxel diffusion heterogeneities in time and space, respectively, and are believed to depend on diffusion times. Studies on this time dependency are limited, largely because the diffusion time cannot vary over a board range in a conventional spin-echo echo-planar imaging sequence due to the accompanying T2 decays. In this study, we investigated the time-dependency of the CTRW model in Sephadex gel phantoms across a broad diffusion time range by employing oscillating-gradient spin-echo, pulsed-gradient spin-echo, and pulsed-gradient stimulated echo sequences. We also performed Monte Carlo simulations to help understand our experimental results. It was observed that the diffusion process fell into the Gaussian regime at extremely short diffusion times whereas it exhibited a strong time dependency in the CTRW parameters at longer diffusion times.