Electric Cell-substrate Impedance Sensing (ECIS) is an impedance-based, real-time, and label-free measuring system for monitoring cellular activities in tissue culture. Previously, ECIS wound healing assay has been used to wound cells with high electric current and monitor the subsequent cell migration. In this study, we applied ECIS electric fence (EF) method, an alternative to electrical wounding, to assess the effects of different surface coatings on human keratinocyte (HaCaT) migration. The EF prevents inoculated cells from attaching or migrating to the fenced electrode surface while maintaining the integrity of the surface coating. After the EF is turned off, cells migrate into the cell-free area, and the increase in measured impedance is monitored. We cultured HaCaT cells on gold electrodes without coating or coated with poly-L-lysin (PLL), poly-D-lysine (PDL), or type-I collagen. We quantified migration rates according to the different slopes in the impedance time series. It was observed that either poly-L-lysine (PLL) or poly-D-lysine (PDL) limits cell adhesion and migration rates. Furthermore, the surface charge of the coated substrate in the culture condition positively correlates with the cell adhesion and migration process. Our results indicate that the EF method is useful for determining cell migration rates on specific surface coatings.
Electric cell–substrate impedance sensing (ECIS) has been used as a real-time impedance-based method to quantify cell behavior in tissue culture. The method is capable of measuring both the resistance and capacitance of a cell-covered microelectrode at various AC frequencies. In this study, we demonstrate the application of high-frequency capacitance measurement (f = 40 or 64 kHz) for the sensitive detection of both the micromotion and wound-healing migration of human mesenchymal stem cells (hMSCs). Impedance measurements of cell-covered electrodes upon the challenge of various concentrations of carbonyl cyanide 4-(trifluoromethoxy)phenylhydrazone (FCCP), from 0.1 to 30 μM, were conducted using ECIS. FCCP is an uncoupler of mitochondrial oxidative phosphorylation (OXPHOS), thereby reducing mitochondrial ATP production. By numerically analyzing the time-series capacitance data, a dose-dependent decrease in hMSC micromotion and wound-healing migration was observed, and the effect was significantly detected at levels as low as 0.1 μM. While most reported works with ECIS use the resistance/impedance time series, our results suggest the potential use of high-frequency capacitance time series for assessing migratory cell behavior such as micromotion and wound-healing migration.