There is an increasing need for safe and simple techniques for sensing devices and prostheses implanted inside the human body. Microwave wireless inspection may be an appropriate technique for it. The implanted device may have specific characteristics that allow to distinguish it from its environment. A new sensing technique based on the principle of differential resonance is proposed and its basic parameters are discussed. This technique allows to use the implant as a signal scattering device and to detect changes produced in the implant based on the corresponding change in its scattering signature. The technique is first tested with a canonic human phantom and then applied to a real in vivo clinical experiment to detect coronary stents implanted in swine animals.
Electrocardiogram (ECG) synchronization is useful to avoid the effects of cardiac motion in medical measurements, and is widely used in standard medical imaging. A number of medical equipment include embedded commercial synchronizers. However, the use of independent synchronization modules is sometimes needed when several non-integrated instruments are used, or in the development of new medical instruments and procedures. We present a simple low-cost ECG synchronizer module based on an Arduino controller board that converts the ECG signal into a transistor-transistor-logic (TTL) one, allowing real-time medical measurements triggered at specific phases of the cardiac cycle. The device and conversion algorithm developed is optimized in vitro using synthetic and human ECG signals, and tested in vivo on three swine specimens. Error rates during the in vivo testing stage remain below the 2% of the cycles in all animals and critical false positives are less than 1%, which is sufficient for most applications. Possible algorithm updates are discussed if its performance needs to be improved.