The American Academy of Pediatrics (AAP) has endorsed Critical Congenital Heart Disease (CCHD) screening using pulse oximetry nationwide, but, however, acknowledges that altitude may impact failure rates and alternative algorithms may be required at high altitudes. We therefore evaluated a modified screening protocol at an altitude of 6200 feet with the hypothesis that modifications could decrease failure rates. We evaluated 2001 well, newborn infants ≥35 weeks gestation using a modified protocol, which included a lower saturation cutoff for the first screen (85% instead of the AAP recommended 90%) and an oxygen hood intervention between the first two screens. Using our modified screening algorithm, we found a 0.3% failure rate, which was similar to the 0.2% sea-level rate and statistically different from the 1.1% rate identified in a recent study at similar altitude. Had the AAP protocol been used, the failure rate would have increased to 0.8%, which is similar to prior reports near this altitude. Echocardiograms were performed on failing newborns with no CCHD identified. A Birth Defects Registry Database review demonstrated one newborn with CCHD was missed after meeting AAP passing criteria. Overall, this study demonstrates that an alternative algorithm can be implemented at moderate altitude with decreased failure rate and comparable false negative rate.
Pulse oximetry screening for critical congenital heart disease (CCHD) has been recommended by the American Academy of Pediatrics (AAP). The objectives of this study are to describe saturation data, and to evaluate the effectiveness of AAP-recommended pulse oximetry screening guidelines applied retrospectively to a cohort of newborns with known CCHD at moderate altitude (5557 feet, Aurora, Colorado). Data related to seven critical congenital heart disease diagnoses were extracted from electronic health records (pulse oximetry, prostaglandin administration, and oxygen supplementation). Descriptive epidemiologic data were calculated. 158 subjects were included in this analysis; the AAP pulse oximetry screening protocol was applied to 149 subjects. Mean pre-ductal and post-ductal pulse oximetry values of the infants known to have CCHD at 24 h of life were 87.1% ± 7.2 and 87.8% ± 6.3, respectively. Infants treated with prostaglandins and oxygen had lower oximetry readings. The screening algorithm would have identified 80.5% of infants with known CCHDs (120/149 subjects). Additionally, sequential pulse oximetry screening based on the AAP-recommended protocol was able to identify a true positive screen capture rate of 80.5% at moderate altitude.