We investigated imaging conditions for the distribution of lutetium oxodotreotide (Lu-177 DOTATATE) in the body during peptide receptor radionuclide therapy for neuroendocrine tumor (NET). We investigated imaging conditions using gamma rays emitted from the radionuclide. The gamma rays had energy peaks at 113 and 208 keV and characteristic X-rays at 56 keV. Image quality was compared by utilizing a combination of low–medium-energy general-purpose (LMEGP) and medium-energy general-purpose (MEGP) collimators. This study included the measurement of total spatial resolution (Full Width at Half Maximum) using a line source phantom. We compared the image quality of static images using a plane phantom and SPECT images using a cylindrical phantom. This comparison involved assessing recovery coefficient curves, count ratio, and %CV. Imaging evaluation was also performed on one NET patient. In phantom studies and the clinical study, comparing the combination of the three energy peaks (56 + 113 + 208 keV) using the LMEGP collimator and the conventional combination (113 + 208 keV) using the MEGP collimator revealed a count ratio of 1.9 times the maximum, stable %CV, and the best image quality.
The step-and-shoot (SS) mode and continuous mode are currently used for single-photon emission computed tomography (SPECT) scan mode, and a new scan mode that combines both modes, step-and-shoot plus continuous (SSC) mode, was developed. It is expected to allow a shorter scan time and lower injected dose because the SSC mode is more sensitive than the SS mode. We confirmed the image quality of this scan mode, including various quantitative correction methods for scatter (SC), attenuation (AC), and resolution recovery (RR) in a phantom study and clinical case study. Image quality was evaluated by the count, contrast-to-noise ratio (CNR), and percent of the coefficient of variation (%CV). Independent of the correction methods, the count, CNR, and %CV of the SSC mode were superior to those of the SS mode. The ACSCRR was the best method, with a maximum increased rate of 66.4% in counts and 57.8% in CNR for the 13-mm sphere and 19.6% in CNR for other sphere sizes. The %CV for the SSC mode was the best for AC and ACRR, which was at 15.1%. With regards to attaining short bone SPECT scan time, the combination of the SSC mode and ACRR or ACSCRR demonstrated the best physical performance.
Recently, the use of saliency maps to evaluate the image quality of nuclear medicine images has been reported. However, that study only compared qualitative visual evaluations and did not perform a quantitative assessment. The study’s aim was to demonstrate the possibility of using saliency maps (calculated from intensity and flicker) to assess nuclear medicine image quality by comparison with the evaluator’s gaze data obtained from an eye-tracking device. We created 972 positron emission tomography images by changing the position of the hot sphere, imaging time, and number of iterations in the iterative reconstructions. Pearson’s correlation coefficient between the saliency map calculated from each image and the evaluator’s gaze data during image presentation was calculated. A strong correlation (r ≥ 0.94) was observed between the saliency map (intensity) and the evaluator’s gaze data. This trend was also observed in images obtained from a clinical device. For short acquisition times, the gaze to the hot sphere position was higher for images with fewer iterations during the iterative reconstruction. However, no differences in iterations were found when the acquisition time increased. Saliency by flicker could be applied to clinical images without preprocessing, although compared with the gaze image, it increased slowly.