We investigated the relationships between radiation dose indices and body size indices in adult body computed tomography (CT). A total of 3200 CT scans of the thoracic, abdominal, abdominopelvic, or thoraco-abdominopelvic regions performed using one of four CT scanners were analyzed. Volume CT dose index (CTDIvol) and dose length product (DLP) were compared with various body size indices derived from CT images (water-equivalent diameter, WED; effective diameter, ED) and physical measurements (weight, weight/height, body mass index, and body surface area). CTDIvol showed excellent positive linear correlations with WED and ED. CTDIvol also showed high linear correlations with physical measurement-based indices, whereas the correlation coefficients were lower than for WED and ED. Among the physical measurement-based indices, weight/height showed the strongest correlations, followed by weight. Compared to CTDIvol, the correlation coefficients with DLP tended to be lower for WED, ED, and weight/height and higher for weight. The standard CTDIvol values at 60 kg and dose increase ratios with increasing weight, estimated using the regression equations, differed among scanners. Radiation dose indices closely correlated with body size indices such as WED, ED, weight/height, and weight. The relationships between dose and body size differed among scanners, indicating the significance of dose management considering body size.
The diagnostic reference levels (DRLs) for pediatric brain computed tomography (CT) are provided for groups divided according to age. We investigated the relationships of radiation dose indices (volume CT dose index and dose length product) with age and weight, as continuous variables, in pediatric brain CT. In a retrospective analysis, 980 pediatric brain CT examinations were analyzed. Curve fitting was performed for plots of the CT dose indices versus age and weight, and equations to estimate age- and weight-dependent standard dose indices were derived. Standard dose indices were estimated using the equations, and the errors were calculated. The results showed a biphasic increase in dose indices with increasing age and weight, characterized by a rapid initial and subsequent slow increase. Logarithmic, power, and bilinear functions were well fitted to the plots, allowing estimation of standard dose indices at an arbitrary age or weight. Error analysis suggested that weight was mildly better than age and that the best results were obtained with the bilinear function. Curve fitting of the relationship between CT dose indices and age or weight facilitates the determination of standard dose indices in pediatric brain CT at each facility and is expected to aid the establishment and application of the DRLs.
We investigated the relationship between the head size and radiation dose in pediatric brain computed tomography (CT) to evaluate the validity of automatic exposure control (AEC). Phantom experiments were performed to assess image noise with and without AEC, and indicated that AEC decreased differences in noise between slices of different section sizes. Retrospective analysis was conducted on 980 pediatric brain CT scans where the tube current was determined using AEC. The water equivalent diameter (WED) was employed as an index of the head size, and mean WED for each image set (WEDmean) and WED for each slice (WEDslice) were used for analysis. For the image-set-based analysis, volume CT dose index (CTDIvol) was compared to WEDmean. For the slice-based analysis, the tube current was compared to WEDslice using 20 of the 980 sets. Additionally, CTDIvol and WEDmean were compared between male and female patients matched for age, weight, or WEDmean. CTDIvol increased with increasing WEDmean, and an exponential curve was closely fitted to the relationship. Tube current changed similarly to the change in WEDslice for each image set, and an exponential curve was well-fitted to the plots of tube current against WEDslice when data from the 20 sets were pooled together. Although CTDIvol and WEDmean were slightly but significantly larger for male than female patients after matching for age or weight, a sex-dependent difference in CTDIvol was not found after matching for WEDmean. This study indicated successful dose modulation using AEC according to the head size for each patient and each slice location. The application of AEC to pediatric brain CT is recommended for radiation dose optimization.
The effective dose (ED) in computed tomography (CT) may be calculated by multiplying the dose–length product (DLP) by a conversion factor. As children grow, automatic exposure control increases the DLP, while the conversion factor decreases; these two changes affect the ED in opposite ways. The aim of this study was to investigate the methods of ED estimation according to age in pediatric brain CT. We retrospectively analyzed 980 brain CT scans performed for various clinical indications in children. The conversion factor at each age, in integer years, was determined based on the values at 0, 1, 5, and 10 years provided by the International Commission on Radiological Protection (ICRP), using a curve (curve method) or lines (linear method). In the simple method, the ED was estimated using the ICRP conversion factor for the closest age. We also analyzed the ED estimated by a radiation dose management system. Although the median DLP at each age increased with age, the median ED estimated by the curve method was highest at 0 years, decreased with age, and then plateaued at 9 years. The linear method yielded mildly different results, especially at 2 and 3 years. The ED estimated by the simple method or the radiation dose management system showed inconsistent, up-and-down changes with age. In conclusion, the ED in pediatric brain CT decreases with age despite increased DLP. Determination of the conversion factor at each age using a curve is expected to contribute to estimating the ED in pediatric CT according to age.
Estimation of the standard radiation dose at each imaging facility is required for radiation dose management, including establishment and utilization of the diagnostic reference levels. We investigated methods to estimate the standard dose for pediatric brain computed tomography (CT) using a small number of data. From 980 pediatric brain CT examinations, 25, 50, and 100 examinations were randomly extracted to create small, medium, and large datasets, respectively. The standard dose was estimated by applying grouping and curve-fitting methods for 20 datasets of each sample size. For the grouping method, data were divided into groups according to age or body weight, and the standard dose was defined as a median value in each group. For the curve-fitting methods, logarithmic, power, and bilinear functions were fitted to plots of radiation dose against age or weight, and the standard dose was calculated at the designated age or weight using the derived equation. When the sample size was smaller, the random variations of the estimated standard dose were larger. Better estimation of the standard dose was achieved with the curve-fitting methods than with the grouping method. Power fitting appeared to be more effective than logarithmic and bilinear fittings for suppressing random variation. Determination of the standard dose for pediatric brain CT by the curve-fitting method is recommended to improve radiation dose optimization at facilities performing the imaging procedure infrequently.
Background: In computed tomography (CT), automatic exposure control (AEC) determines the tube current and thus the radiation dose based on scout images. We investigated CT dose modulation using two versions of CARE Dose 4D, Siemens AEC software. Methods: A cylindrical phantom and an anthropomorphic phantom with the upper extremities raised or down were imaged. The CT tube current was determined using two versions of CARE Dose 4D and different scout directions: the posteroanterior scout image alone (PA scout), the lateral scout image alone (Lat scout), and the combination of the PA and Lat scout images (PA + Lat scout). The new version is designed to utilize the Lat image solely for off-center correction when both PA and Lat images are available. Experiments were performed at various vertical positions and with various scout imaging parameters. Results: The influence of the scout direction on CT dose was demonstrated, with variations depending on the imaging object and software version. The CT dose determined with the PA scout varied according to vertical positioning, presumably due to changes in image magnification. Such effects were small with the Lat scout or PA + Lat scout. Decreasing the tube voltage or tube current in scout imaging affected CT dose modulation with the Lat scout but not with the PA scout. With the PA + Lat scout, the effects of scout parameters were evident using the previous version but minimal using the new version. Conclusions: Off-center correction in the new version functioned appropriately. Because the behavior of an AEC system is complicated, it is recommended to examine the characteristics of each AEC system under various imaging conditions.