Longfu Jiang

Dr. Longfu Jiang

orcid

Research Focus & Metrics

Academic Career Length

6Years (2021-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
30 June 2022 in Polish Heart Journal
Verified
Journal Article
Closed access
PDF file
HWHao Wu
Hao Wu
LJLongfu Jiang
Longfu Jiang
JSJiabo Shen
Jiabo Shen
The present study aimed to evaluate the feasibility and safety of the novel left bundle branch pacing (LBBP) procedure that uses isoelectric interval as an endpoint for lead implantation. A total of 41 patients with indications for pacing were enrolled. All patients underwent a novel LBBP procedure guided by recording an isoelectric interval as an endpoint for lead implantation. The procedural details and electrophysiological characteristics were then analyzed. A total of 38/41 (92.7%) cases were confirmed of left bundle branch (LBB) capture. An isoelectric interval was observed in 36/41 cases (87.8%). A total of 36/41 (87.8%) cases with LBB potential were observed. The mean unipolar LBBP threshold at the implant was 0.5 ± 0.2 V. The mean sensed amplitude of the R wave and the pacing impedance at the implant were 12.9 ± 5.0 mV and 723.5 ± 117.1 Ω. During the final threshold testing, a transition from non-selective to selective LBBP (S-LBBP) was demonstrated in 26 patients. A transition from non-selective LBBP (NS-LBBP) to left ventricular septal myocardial capture was observed in 12 patients. Using an isoelectric interval as an endpoint to guide the LBBP was feasible in a high proportion of captured LBB cases.

8 March 2022 in Canadian Journal of Cardiology
Verified
Journal Article
Closed access
JSJiabo Shen
Jiabo Shen
LJLongfu Jiang
Longfu Jiang
XCXiaojie Cai
Xiaojie Cai
HWHao Wu
Hao Wu
LPLifang Pan
Lifang Pan

17 June 2022 in Heart Rhythm
Verified
Journal Article
Closed access
HWHao Wu
Hao Wu
LJLongfu Jiang
Longfu Jiang
JSJiabo Shen
Jiabo Shen
LZLu Zhang
Lu Zhang
JZJinyan Zhong
Jinyan Zhong
SZShanshan Zhuo
Shanshan Zhuo

14 September 2022 in Canadian Journal of Cardiology
Verified
Journal Article
Closed access
JSJiabo Shen
Jiabo Shen
LJLongfu Jiang
Longfu Jiang
HWHao Wu
Hao Wu
XCXiaojie Cai
Xiaojie Cai
SZShanshan Zhuo
Shanshan Zhuo
LPLifang Pan
Lifang Pan

11 January 2023 in Frontiers in Cardiovascular Medicine
Verified
Journal Article
Open access
JSJiabo Shen
Jiabo Shen
LJLongfu Jiang
Longfu Jiang
HWHao Wu
Hao Wu
HLHengdong Li
Hengdong Li
LZLu Zhang
Lu Zhang
LPLifang Pan
Lifang Pan
Objective: The characteristics of discrete intracardiac electrogram (EGM) in selective left bundle branch (SLBB) pacing (SLBBP) have not been described in detail previously. This study aimed to examine the effect of different high-pass filter (HPF) settings on discrete local ventricular components in an intracardiac EGM and to analyze its possible mechanisms.Methods: This study included 144 patients with indications of permanent cardiac pacing. EGMs were collected at four different HPF settings (30, 60, 100, and 200 Hz) with a low-pass filter at 500 Hz, and their possible mechanisms were analyzed.Results: LBBP was successfully achieved in 91.0% (131/144) of patients. SLBBP was achieved in 123 patients. The occurrence rates of discrete local ventricular EGM were 16.7, 33.3, 72.9, and 85.4% for HPF settings of 30, 60, 100, and 200 Hz, respectively. The analysis of discrete EGM detection showed significant differences between the different HPF settings. By using the discrete local ventricular component and isoelectric interval as the SLBB capture golden standard, the results of EGMs revealed that the 30 Hz HPF has a sensitivity of 19% and specificity of 100%. The 60 Hz HPF had a sensitivity of 39% and a specificity of 100%. The 100 Hz HPF had a sensitivity of 85% and a specificity of 100%. The 200 Hz HPF had a sensitivity of 100% and specificity of 100%.Conclusion: An optimal HPF setting of 200 Hz is recommended for discrete local ventricular EGM detection. A discrete local ventricular EGM should exhibit an isoelectric interval. A steep deflection and high-frequency ventricular EGM morphology nearly identify an intrinsic EGM morphology.

6 November 2024 in Heart Rhythm
Verified
Journal Article
Open access
JSJiabo Shen
Jiabo Shen
LJLongfu Jiang
Longfu Jiang
HWHao Wu
Hao Wu
LZLu Zhang
Lu Zhang
HLHengdong Li
Hengdong Li
LPLifang Pan
Lifang Pan
The interrupted technique of left bundle branch pacing (LBBP) limits the continuous monitoring of paced electrocardiogram (ECG) and intracardiac electrogram (EGM) transitions, which may result in overlooked or misinterpreted subtle transitions. This study aimed to explore the electrophysiological characteristics of lead position-dependent EGM continuous transition to evaluate lead depth and investigate the clinical significance of transseptal pacing modalities. A continuous pacing and recording technique enabled by a rotatable connector was employed to allow the real-time monitoring of progressive changes in paced EGM and ECG morphology. Careful observations were conducted to evaluate whether there were significant changes in the amplitude and morphology of the ventricular current of injury (COI), R-wave peak times of V1 and V6, QRS duration, and impedance at different interventricular septal depths. The study included 105 patients. Nonselective LBBP was achieved in 94 (89.5 %) patients, of whom 88 (83.8%) achieved selective LBBP (SLBBP). Left ventricular septal pacing was confirmed in 11 (11.5 %) patients. The amplitude of ventricular EGM predictably changed with radial septum depth and peaked in the interventricular septum (26.3±11.3 mV). As the lead was inserted into the left ventricular subendocardium, the ventricular COI declined to a level approximating that of the right septum (11.7 ± 6.3 mV for SLBBP versus 10.4 ± 5.8 mV for right ventricular septal pacing). When selective LBB capture occurred, significant morphological transitions in the ventricular COI were observed in the unfiltered EGM. The continuous recording technique provides a more detailed understanding of pacing lead radial depth throughout implantation. COI amplitude and morphology variations can identify different pacing modalities, particularly in recognizing SLBBP.