The use of finite element analysis (FEA) has increased rapidly over the last decennia and has become a popular tool to design implants, osteosynthesis plates and prostheses. With increasing computer capacity and the availability of software applications, it has become easier to employ the FEA. However, there seems to be no consensus on the input variables that should be applied to representative FEA models of the human mandible. This review aims to find a consensus on how to define the representative input factors for a FEA model of the human mandible. A literature search carried out in the PubMed and Embase database resulted in 137 matches. Seven papers were included in this current study. Within the search results, only a few FEA models had been validated. The material properties and FEA approaches varied considerably, and the available validations are not strong enough for a general consensus. Further validations are required, preferably using the same measuring workflow to obtain insight into the broad array of mandibular variations. A lot of work is still required to establish validated FEA settings and to prevent assumptions when it comes to FEA applications.
1 January 2019 in International Journal of Oral and Maxillofacial Surgery
Zygomatic implants are used in patients with maxillary defects to improve the retention and stability of obturator prostheses, thereby securing good oral function. Prosthetic-driven placement of zygomatic implants is even difficult for experienced surgeons, and with a free-hand approach, deviation from the preplanned implant positions is inevitable, thereby impeding immediate implant-retained obturation. A novel, digitalized workflow of surgical planning was used in 10 patients. Maxillectomy was performed with 3D-printed cutting, and drill guides were used for subsequent placement of zygomatic implants with immediate placement of implant-retained obturator prosthesis. The outcome parameters were the accuracy of implant positioning and the prosthetic fit of the obturator prosthesis in this one-stage procedure. Zygomatic implants (n = 28) were placed with good accuracy (mean deviation 1.73 ± 0.57 mm and 2.97 ± 1.38° 3D angle deviation), and in all cases, the obturator prosthesis fitted as pre-operatively planned. The 3D accuracy of the abutment positions was 1.58 ± 1.66 mm. The accuracy of the abutment position in the occlusal plane was 2.21 ± 1.33 mm, with a height accuracy of 1.32 ± 1.57 mm. This feasibility study shows that the application of these novel designed 3D-printed surgical guides results in predictable zygomatic implant placement and provides the possibility of immediate prosthetic rehabilitation in head and neck oncology patients after maxillectomy.
Objectives: In case of surgical removal of oral squamous cell carcinomas, a resection of mandibular bone is frequently part of the treatment. Nowadays, such resections frequently include the application of 3D virtual surgical planning (VSP) and guided surgery techniques. In this paper, current methods for 3D VSP leads for optimisation of the workflow, and patient‐specific application of guides and implants are reviewed.Recent findings: Current methods for 3D VSP enable multi‐modality fusion of images. This fusion of images is not restricted to a specific software package or workflow. New strategies for 3D VSP in Oral and Maxillofacial Surgery include finite element analysis, deep learning and advanced augmented reality techniques. These strategies aim to improve the treatment in terms of accuracy, predictability and safety.Conclusions: Application of the discussed novel technologies and strategies will improve the accuracy and safety of mandibular resection and reconstruction planning. Accurate, easy‐to‐use, safe and efficient three‐dimensional VSP can be applied for every patient with malignancies needing resection of the mandible.
Introduction: When the application of a free vascularised flap is not possible, a segmental mandibular defect is often reconstructed using a conventional reconstruction plate. Mechanical failure of such reconstructions is mostly caused by plate fracture and screw pull‐out. This study aims to develop a reliable, mechanically superior, yet slender patient‐specific reconstruction plate that reduces failure due to these causes.Patients and Methods: Eight patients were included in the study. Indications were as follows: fractured reconstruction plate (2), loosened screws (1) and primary reconstruction of a mandibular continuity defect (5). Failed conventional reconstructions were studied using finite element analysis (FEA). A 3D virtual surgical plan (3D‐VSP) with a novel patient‐specific (PS) titanium plate was developed for each patient. Postoperative CBCT scanning was performed to validate reconstruction accuracy.Results: All PS plates were placed accurately according to the 3D‐VSP. Mean 3D screw entry point deviation was 1.54 mm (SD: 0.85, R: 0.10–3.19), and mean screw angular deviation was 5.76° (SD: 3.27, R: 1.26–16.62). FEA indicated decreased stress and screw pull‐out inducing forces. No mechanical failures appeared (mean follow‐up: 16 months, R: 7–29).Conclusion: Reconstructing mandibular continuity defects with bookshelf‐reconstruction plates with FEA underpinning the design seems to reduce the risk of screw pull‐out and plate fractures.
1 May 2018 in Journal of Cranio-Maxillofacial Surgery