A method for the simultaneous determination of alendronate, pamidronate, ibandronate and risedronate using ion chromatography with integrated pulsed amperometric detection (IPAD) has been developed. The electrochemical behavior showed the catalytic currents of these bisphosphonates are based on the oxidation of amines in their structures. Because the bisphosphonates are polar compounds and present as anions under alkaline condition, therefore, they can be separated by anion exchange chromatography. A Dionex AS18 column (250 mm x 2 mm) and an AG18 column (50 mm x 2 mm) and 24 mmol/L NaOH solution were used for the separation. Multi-step potential waveform parameters were optimized to maximize the signal-to-noise ratio (S/N), which exhibited adsorption/desorption catalytic features at the gold electrode surface in alkaline solution. Utilizing the optimized waveform, the method showed good linearity (r2 = 0.9972 - 0.9995), satisfactory repeatability (relative standard deviations (RSDs) of the peak areas in the range of 0.84% - 1.37%) and sufficient sensitivity (limits of detection of 0.061 - 0.18 microg/mL) for the identification of the four bisphosphonates. The recoveries were 80.81% - 97.32% with the RSDs of 1.46% - 3.02%. It is demonstrated that this method is a rapid and simple one for the determination of the four bisphosphonates in human plasma.
An ion chromatograph single-pump column-switching system was developed for the determination of ammonium in high salt matrices. The simple system included a pump, a suppressor, two valves, two columns, a single eluent, and a conductivity detector. Using this method, both online collection and matrix elimination were achieved. Ammonium was eluted from the acceptor loop to the analytical column circularly. Under the optimized separation conditions, the method showed good linearity (r = 0.9997) in the range of 0.05‐2.0 mg/L and satisfactory repeatability (RSD < 4%, n = 6). The limit of detection was 4 μg/L (S/N = 3) with 25 μl injection. The utilization of the method was demonstrated by the analysis of seawater, with satisfactory spiked recoveries of ammonium between 98.8% and 108.7%. The result showed that the single-pump column-switching system was convenient and practical for the determination of low-level ammonium in high-salinity samples.