This paper presents a study regarding the design and the experimental setup of a medical robotic system for brachytherapy using tribology analysis. The robotic system is composed of a collaborative robotic arm and a multi-needle brachytherapy instrument controlled using a unified control system embedding a haptic device and force-feedback. This work is oriented towards identifying the technical characteristics of the system components to determine the accuracy of the procedure, as well as using different scenarios for needle insertion in ex vivo porcine liver tissue in order to determine the forces required for insertion and extraction of the needle and the friction coefficient that accompanies the previously mentioned forces. Subsequent to the computation of the friction forces, the normal forces and the wear during the needle insertion are determined with the scope of predicting the lifecycle of some components of the medical device.
Remote center-of-motion mechanisms are essential in minimally invasive surgery because they allow surgical instruments or an endoscopic camera to pivot around a trocar entry point while eliminating lateral motion at the incision. This paper presents the design, kinematic modeling, prototype implementation and preliminary evaluation under laboratory conditions of a compact, spherical, remote center-of-motion robot for minimally invasive surgical orientation tasks. The proposed mechanism uses a spherical kinematic architecture actuated by a contra-rotating differential gearbox. This gearbox generates two coaxial output rotations of equal magnitude and opposite direction from a single input, mechanically synchronizing the opposed motion of the two base links and eliminating the need for cable-pulley transmission or dual electronically synchronized motors. A second actuator chain rotates the gearbox assembly around the base axis, thereby decoupling the extension–retraction motion from base-axis rotation. Forward and inverse kinematic formulations were derived for teleoperation of the robot using a 7 degrees of freedom haptic device and for remote center-of-motion orientation control using a 3-axis joystick. A proof-of-concept prototype was developed and integrated with a custom embedded controller, closed-loop motor control, a master-console interface and video feedback loop. The system was evaluated in a phantom-torso setup using a custom endoscopic camera, internal visual markers and an OptiTrack-based measurement of the remote center-of-motion accuracy. The qualitative experiment confirmed functional integration of the mechanical, electronic and software subsystems, while the optical-tracking measurement showed that the pivot constraint was maintained with a mean deviation of 1.69 mm and a root-mean-square deviation of 2.13 mm over the analyzed orientation sweep. The main limitations remain the 1:1 gearbox ratio, limited actuator torque, additively manufactured gearing and the absence of repeated-trial repeatability and full workspace characterization.