Objective:Evaluation of absolute radiation exposure values for interventional radiologists (IRs) using a multiaxis interventional flat-panel C-arm cone beam CT (CBCT) system with three-dimensional laser guidance for biopsy in a triple-modality, abdominal phantom.Methods:In the phantom, eight lesions were punctured in two different angles (in- and out-of-plane) using CBCT. One C-arm CT scan was performed to plan the intervention and one for post-procedural evaluation. Thermoluminescent dosemeters (TLDs) were used for dose measurement at the level of the eye lens, umbilicus and ankles on a pole representing the IRs. All measurements were performed without any lead protection. In addition, the dose–area product (DAP) and air kerma at the skin entrance point was documented.Results:Mean radiation values of all TLDs were 190 µSv for CBCT (eye lens: 180 µS, umbilicus: 230 µSv, ankle: 150 µSv) without a significant difference (p > 0.005) between in- and out-of-plane biopsies. In terms of radiation exposure of the...