The accuracies of various navigation systems in total hip arthroplasty (THA) have been described; however, the accuracy of cup orientation with an accelerometer-based navigation system has not been reported. The purpose of this study was to compare the accuracies of computed tomography (CT)- and accelerometer-based navigation systems for cup orientation in THA. In this prospective study, 30 patients who underwent cementless THA via anterolateral approach in the lateral decubitus position were analysed. A CT-based navigation system (30 hips) and an accelerometer-based navigation system (30 hips) were used simultaneously. The accuracy of cup orientation (absolute difference between intraoperative vs. postoperative measurements) was compared between the navigation systems using postoperative CT. The accuracy of cup inclination was 3.2 ± 2.4° in the CT-based navigation group and 4.1 ± 3.7° in the accelerometer-based navigation group (p = 0.3035). The accuracy of cup anteversion was 3.0° ± 2.5° in the CT-based navigation group and 6.8° ± 4.8° in the accelerometer-based navigation group. Cup anteversion was significantly more accurate with the CT-based navigation system than with accelerometer-based navigation (p = 0.0009). Multiple regression analysis demonstrated that the malposition in cup anteversion was positively correlated with the change in pelvic tilt and loosening of the reference antenna. Although, these results are only true for this specific accelerometer system, cup positioning was significantly more accurate with the CT-based navigation system than an accelerometer-based navigation in the lateral decubitus position. This is because of considerable discrepancies in the sagittal pelvic tilt, resulting in variability in cup anteversion angle with the use of an accelerometer-based navigation system.
Background: Inadequate acetabular component orientation is associated with postoperative impingement, dislocation, and accelerated polyethylene wear. Computed tomography (CT)-based navigation systems provide accuracy for total hip arthroplasty (THA) but are not available in all facilities. Accelerometer-based navigation systems are inexpensive, but their accuracy remains undetermined. This study compares the accuracy of cup orientation in THA using CT-based and accelerometer-based navigation systems. Methods: This retrospective study included 35 consecutive patients (11 males, 24 females; mean age, 65 years) who underwent primary cementless THA via an anterolateral approach in the supine position. Both CT-based and accelerometer-based navigation systems were used simultaneously. The accuracy of cup orientation was compared between the two systems using postoperative CT. Results: The accuracy of cup inclination was 2.7° ± 2.0° in the CT-based group and 3.3° ± 2.4° in the accelerometer-based group. The accuracy of cup anteversion was 2.8° ± 2.6° in the CT-based group and 3.4° ± 2.2° in the accelerometer-based group. No significant difference was observed in cup inclination (p = 0.29) or cup anteversion (p = 0.34) between CT-based and accelerometer-based navigation. Conclusions: The accuracy of cup positioning did not differ significantly between CT-based and accelerometer-based navigation systems.