Multiplexed bioimaging systems have triggered the development of effective assays, contributing new biological information. Although electrochemical imaging is beneficial for quantitative analysis in real time, monitoring multiple cell functions is difficult. We have developed a novel electrochemical imaging system, herein, using a large-scale integration (LSI)-based amperometric device for detecting multiple biomolecules simultaneously. This system is designated as an electrochemicolor imaging system in which the current signals from two different types of biomolecules are depicted as a multicolor electrochemical image. The mode-selectable function of the 400-electrode device enables the imaging system and two different potentials can be independently applied to the selected electrodes. The imaging system is successfully applied for detecting multiple cell functions of the embryonic stem (ES) cell and the rat pheochromocytoma (PC12) cell aggregates. To the best of our knowledge, this is the first time that a real-time electrochemical mapping technique for multiple electroactive species, simultaneously, has been reported. The imaging system is a promising bioanalytical method for exploring complex biological phenomena.
An electrochemical device, which consists of electrode arrays, nanocavities, and microwells, was developed for multi-electrochemical detection with high sensitivity. A local redox cycling-based electrochemical (LRC-EC) system was used for multi-electrochemical detection and signal amplification. The LRC-EC system consists of n2 sensors with only 2n bonding pads for external connection. The nanocavities fabricated in the sensor microwells enable significant improvement of the signal amplification compared with the previous devices we have developed. The present device was successfully applied for evaluation of embryoid bodies (EBs) from embryonic stem (ES) cells via electrochemical measurements of alkaline phosphatase (ALP) activity in the EBs. In addition, the EBs were successfully trapped in the sensor microwells of the device using dielectrophoresis (DEP) manipulation, which led to high-throughput cell analysis. This device is considered to be useful for multi-electrochemical detection and imaging for bioassays including cell analysis.