Computed tomography examinations have caused high radiation doses for patients, especially for CT scans of the brain. This study aimed to optimize the radiation dose and image quality in adult brain CT protocols. Images were acquired using a Catphan 700 phantom. Radiation doses were recorded as CTDIvol and dose length product (DLP). CT brain protocols were optimized by varying parameters such as kVp, mAs, signal-to-noise ratio (SNR) level, and Clearview iterative reconstruction (IR). The image quality was also evaluated using AutoQA Plus v.1.8.7.0 software. CT number accuracy and linearity had a robust positive correlation with the linear attenuation coefficient (µ) and showed more inaccurate CT numbers when using 80 kVp. The modulation transfer function (MTF) showed a higher value in 100 and 120 kVp protocols (p < 0.001), while high-contrast spatial resolution showed a higher value in 80 and 100 kVp protocols (p < 0.001). Low-contrast detectability and the contrast-to-noise ratio (CNR) tended to increase when using high mAs, SNR, and the Clearview IR protocol. Noise decreased when using a high radiation dose and a high percentage of Clearview IR. CTDIvol and DLP were increased with increasing kVp, mAs, and SNR levels, while the increasing percentage of Clearview did not affect the radiation dose. Optimized protocols, including radiation dose and image quality, should be evaluated to preserve diagnostic capability. The recommended parameter settings include kVp set between 100 and 120 kVp, mAs ranging from 200 to 300 mAs, SNR level within the range of 0.7–1.0, and an iterative reconstruction value of 30% Clearview to 60% or higher.
Computed tomography examination plays a vital role in imaging and its use has rapidly increased in radiology diagnosis. This study aimed to assess radiation doses of routine CT protocols of the brain, chest, and abdomen in three different CT scanners, together with a qualitative image quality assessment. Methods: A picture archiving and communication system (PACS) and Radimetrics software version 3.4.2 retrospectively collected patients’ radiation doses. Radiation doses were recorded as the CTDIvol, dose length product, and effective dose. CT images were acquired using the Catphan700 phantom to evaluate image quality. Results: The findings revealed that median values for the CTDIvol and DLP across the brain, chest, and abdomen protocols were lower than the national and international DRLs. Effective doses for brain, chest, and abdomen protocols were also below the median value of R. Smith-Bindman. Neusoft achieved higher spatial frequencies in brain protocols, while Siemens outperformed others in chest protocols. Neusoft consistently exhibited superior high-contrast resolution. Siemens and Neusoft outperformed low-contrast detectability, while Siemens also outperformed the contrast-to-noise ratio. In addition, Siemens had the lowest image noise in brain protocols and high uniformity in chest and abdomen protocols. Neusoft showed the lowest noise in chest and abdomen protocols and high uniformity in the brain protocol. The noise power spectrum revealed that Philips had the highest noise magnitude with different noise textures across protocols and scanners. Conclusions: This study provides a comprehensive evaluation of radiation doses and image quality for three different CT scanners using standard clinical protocols. Almost all CT protocols exhibited radiation doses below the DRLs and demonstrated varying image qualities across each protocol and scanner. Selecting the right CT scanner for each protocol is essential to ensure that the CT images exhibit the best quality among a wide range of CT machines. The MTF, HCR, LCD, CNR, NPS, noise, and uniformity are suitable parameters for evaluating and monitoring image quality.
Background: Computed tomography examinations have produced high radiation doses to patients, especially the CT brain. This study aimed to optimize the radiation dose and image quality in adult CT brain protocol. Materials and Methods: Images were acquired by Catphan 700 phantom. Radiation doses were recorded as CTDIvol and dose length product (DLP). CT brain protocols were optimized by varying parameters such as kVp, mAs, signal-to-noise ratio (SNR) level, and Clearview iterative reconstruction (IR). Image quality were also evaluataion by The AutoQA Plus software. Results: CT Number accuracy and linearity had a robust positive correlation with the linear attenuation coefficient (µ) and showed more inaccurate CT numbers when using 80 kVp. MTF showed a higher value in 100 and 120 kVp protocols, while high contrast spatial resolution showed a higher value in 80 and 100 kVp protocols. Low contrast detectability and CNR tended to increase when using high mAs, SNR and Clearview IR protocol. Noise decreased when using a high radiation dose and a high percentage of Clearview IR. CTDIvol and DLP were increased with increasing kVp, mAs, and SNR levels, while the increasing percentage of Clearview did not affect the radiation dose. Conclusion: Optimized protocols, including radiation dose and image quality, should be evaluated to preserve diagnostic capability. The recommended parameter settings include kVp set between 100-120 kVp, mAs ranging from 200-300 mAs, SNR level within the range of 0.7-1.0, and an iterative reconstruction value of 30% Clearview to 60% or higher.