Blood tests are widely used in modern medicine to diagnose certain illnesses and evaluate the overall health of a patient. To enable testing in resource-limited areas, there has been increasing interest in point-of-care (PoC) testing devices. To process blood samples, liquid mixing with active pumps is usually required, making PoC blood testing expensive and bulky. We explored the possibility of processing approximately 2 μL of whole blood for image flow cytometry using capillary structures that allowed test times of a few minutes without active pumps. Capillary pump structures with five different pillar shapes were simulated using Ansys Fluent to determine which resulted in the fastest whole blood uptake. The simulation results showed a strong influence of the capillary pump pillar shape on the chip filling time. Long and thin structures with a high aspect ratio exhibited faster filling times. Microfluidic chips using the simulated pump design with the most efficient blood uptake were fabricated with polydimethylsiloxane (PDMS) and polyethylene oxide (PEO). The chip filling times were tested with 2 μL of both water and whole blood, resulting in uptake times of 24 s for water and 111 s for blood. The simulated blood plasma results deviated from the experimental filling times by about 35% without accounting for any cell-induced effects. By comparing the flow speed induced by different pump pillar geometries, this study offers insights for the design and optimization of passive microfluidic devices for inhomogenous liquids such as whole blood in sensing applications.
Lateral flow immunoassays (LFIAs) were integrated into microfluidic chips and tested to enhance point-of-care testing (POCT), with the aim of improving sensitivity and expanding the range of CRP detection. The microfluidic approach improves upon traditional methods by precisely controlling fluid speed, thus enhancing sensitivity and accuracy in CRP measurements. The microfluidic approach also enables a one-step detection system, eliminating the need for buffer solution steps and reducing the nitrocellulose (NC) pad area to just the detection test line. This approach minimizes the non-specific binding of conjugated antibodies to unwanted areas of the NC pad, eliminating the need to block those areas, which enhances the sensitivity of detection. The gold nanoparticle method detects CRP in the high-sensitivity range of 1–10 g/mL, which is suitable for chronic disease monitoring. To broaden the CRP detection range, including infection levels beyond 10 g/mL, fluorescent labels were introduced, extending the measuring range from 1 to 70 g/mL. Experimental results demonstrate that integrating microfluidic technology significantly enhances operational efficiency by precisely controlling the flow rate and optimizing the mixing efficiency while reducing fabrication resources by eliminating the need for separate pads, making these methods suitable for resource-limited settings. Microfluidics also provides greater control over fluid dynamics compared to traditional LFIA methods, which contributes to enhanced detection sensitivity even with lower sample volumes and no buffer solution, helping to enhance the usability of POCT. These findings highlight the potential to develop accessible, accurate, and cost-effective diagnostic tools essential for timely medical interventions at the POC.