Drago Resnik

Dr. Drago Resnik


Research Focus & Metrics

Academic Career Length

27Years (2000-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
20 January 2018 in Micromachines
Verified
Journal Article
Open access
PDF file
DRDrago Resnik
Drago Resnik
MMMatej Možek
Matej Možek
BPBorut Pečar
Borut Pečar
Andrej JanežAndrej Janež
Andrej Janež
Vilma Urbančič RovanVilma Urbančič Rovan
Vilma Urbančič Rovan
DVDanilo Vrtačnik
Danilo Vrtačnik
An experimental study of in vivo insulin delivery through microinjection by using hollow silicon microneedle array is presented. A case study was carried out on a healthy human subject in vivo to determine the influence of delivery parameters on drug transfer efficiency. As a microinjection device, a hollow microneedle array (13 × 13 mm2) having 100 microneedles (220 µm high, 130 µm-outer diameter and 50 µm-inner diameter) was designed and fabricated using classical microfabrication techniques. The efficiency of the delivery process was first characterized using methylene blue and a saline solution. Based on these results, the transfer efficiency was found to be predominantly limited by the inability of viable epidermis to absorb and allow higher drug transport toward the capillary-rich region. Two types of fast-acting insulin were used to provide evidence of efficient delivery by hollow MNA to a human subject. By performing blood analyses, infusion of more-concentrated insulin (200 IU/mL, international units (IU)) exhibited similar blood glucose level drop (5–7%) compared to insulin of standard concentration (100 IU/mL), however, significant increase of serum insulin (40–50%) with respect to the preinfusion values was determined. This was additionally confirmed by a distinctive increase of insulin to C-peptide ratio as compared to preinfusion ratio. Moreover, we noticed that this route of administration mimics a multiple dose regimen, able to get a “steady state” for insulin plasma concentration.

1 May 2000 in Microelectronic Engineering
Verified
Journal Article
Closed access
DRDrago Resnik
Drago Resnik
DVD Vrtacnik
D Vrtacnik
UAU Aljancic
U Aljancic
SAS Amon
S Amon

1 September 2015 in Sensors and Actuators A: Physical
Verified
Journal Article
Closed access
TDTine Dolžan
Tine Dolžan
BPBorut Pečar
Borut Pečar
MMMatej Možek
Matej Možek
DRDrago Resnik
Drago Resnik
DVDanilo Vrtačnik
Danilo Vrtačnik

4 March 2013 in Sensors
Verified
Journal Article
Open access
PDF file
BPBorut Pečar
Borut Pečar
DVDanilo Vrtačnik
Danilo Vrtačnik
DRDrago Resnik
Drago Resnik
MMMatej Možek
Matej Možek
UAUroš Aljančič
Uroš Aljančič
DKDejan Križaj
Dejan Križaj
A novel design for a strip-type microthrottle pump with a rectangular actuator geometry is proposed, with more efficient chip surface consumption compared to existing micropumps with circular actuators. Due to the complex structure and operation of the proposed device, determination of detailed structural parameters is essential. Therefore, we developed an advanced, fully coupled 3D electro-fluid-solid mechanics simulation model in COMSOL that includes fluid inertial effects and a hyperelastic model for PDMS and no-slip boundary condition in fluid-wall interface. Numerical simulation resulted in accurate virtual prototyping of the proposed device only after inclusion of all mentioned effects. Here, we provide analysis of device operation at various frequencies which describes the basic pumping effects, role of excitation amplitude and backpressure and provides optimization of critical design parameters such as optimal position and height of the microthrottles. Micropump prototypes were then fabricated and characterized. Measured characteristics proved expected micropump operation, achieving maximal flow-rate 0.43 mL·min−1 and maximal backpressure 12.4 kPa at 300 V excitation. Good agreement between simulation and measurements on fabricated devices confirmed the correctness of the developed simulation model.

26 August 2014 in Microsystem Technologies
Verified
Journal Article
Closed access
PDF file
BPBorut Pečar
Borut Pečar
DVDanilo Vrtačnik
Danilo Vrtačnik
DRDrago Resnik
Drago Resnik
MMMatej Možek
Matej Možek
TDTine Dolžan
Tine Dolžan
DKDejan Križaj
Dejan Križaj

10 April 2019
Verified
Book Chapter
Open access
PDF file
MMMatej Možek
Matej Možek
BPBorut Pečar
Borut Pečar
DRDrago Resnik
Drago Resnik
DVDanilo Vrtačnik
Danilo Vrtačnik
Quality control of piezoelectric micropumps is presented through design, fabrication process, operation, and characterization. The presented study resulted in the extraction of a minimal set of monitored parameters, which is a prerequisite for reliable and stable micropump operation. Micropump fabrication process steps, especially bonding process quality, in correlation with quality control of micropump constituent components (housing, elastomer, and piezoelectric actuator) provided an explanation for deterioration of common micropump characteristics, such as flow vs. backpressure, suction pressure, and excitation signal. These characteristics also manifested in deterioration of other important micropump properties, such as self-priming ability, bubble tolerance, long-term stability, heat dissipation, and temperature operating range. Besides air and DI water pumping, chemical compatibility of constituent materials was confirmed during successful long-term testing of micropumps by pumping media with different viscosity and aggressive media with low pH value. The extracted set of parameters defines input control for micropump fabrication process while at the same time establishes safe operating area of fabricated micropumps. The presented set of parameters provides quality control guidelines and enables a direct comparison from pump-to-pump or run-to-run variations and extraction of influencing design or fabrication parameters.