Raster-scan optoacoustic mesoscopy (RSOM) comes with high potential for in vivo diagnostic imaging in dermatology, since it allows for high resolution imaging of the natural chromophores melanin, and hemoglobin at depths of several millimeters. We have applied ultra-wideband RSOM, in the 10-160 MHz frequency band, to image healthy human skin at distinct locations. We analyzed the anatomical information contained at different frequency ranges of the optoacoustic (photoacoustic) signals in relation to resolving features of different skin layers in vivo. We further compared results obtained from glabrous and hairy skin and identify that frequencies above 60 MHz are necessary for revealing the epidermal thickness, a prerequisite for determining the invasion depth of melanoma in future studies. By imaging a benign nevus we show that the applied RSOM system provides strong contrast of melanin-rich structures. We further identify the spectral bands responsible for imaging the fine structures in the stratum corneum, assessing dermal papillae, and resolving microvascular structures in the horizontal plexus.
We have imaged for the first time to our knowledge human skin in vivo with a raster-scan optoacoustic mesoscopy system based on a spherically focused transducer with a central frequency of 102.8 MHz and large bandwidth (relative bandwidth 105%). Using tissue phantoms we have studied the ability of the system to image vessels of sizes within the anatomically significant range from the key anatomical vasculature sites. The reconstructed images from experiments in vivo show several structures from the capillary loops at the dermal papillae, the horizontal plexus, and the difference between the dermis and the epidermis layers.
Pulsed laser diodes may offer a smaller, less expensive alternative to conventional optoacoustic laser sources; however they do not provide pulse rates faster than a few tens of kHz and emit at wavelengths only within the near-infrared region. We investigated whether continuous wave (CW) laser diodes, which are available in visible and near-infrared regions, can be good optoacoustic light sources when overdriven with a peak current >40-fold higher than the CW absolute maximum. We found that overdriven CW diodes provided ∼10 ns pulses of ∼200 nJ/pulse and repetition rates higher than 600 kHz without being damaged, outperforming many pulsed laser diodes. Using this system, we obtained images of phantoms and mouse ear and human arm in vivo, confirming their use in optoacoustic imaging and sensing.