Achieving accurate and high-sensitivity liquid level detection in medical instruments has always been a knotty task. In this paper, a high-precision, non-contact, flexible capacitive liquid level sensor is proposed, aiming to apply capacitive sensors in test tube liquid level measurement and improving the sensitivity of real-time liquid level sensors. The simulation study is conducted using ANSYS Maxwell and demonstrates the correlation between test tube thickness and sensitivity. A geometric model of the test container and sensing electrodes is established to optimize the design strategy for the physical dimensions of the sensor’s interdigitated (IDT) electrodes based on a flexible printed circuit (FPC). The hardware and software designs are completed based on the FDC2214 capacitive-to-digital converter to collect the capacitance variation data of the sensing electrodes accurately. To assess the system’s performance, an experimental platform for a liquid level sensor system has been constructed, facilitating the measurement, communication, processing, and visualization of liquid levels. The performance results demonstrate that the system is capable of accurately measuring the effective liquid level range within a standard 5 mL test tube with a resolution of up to 1 mm, as well as a sensitivity of 78.68 fF/mm, verifying the simulation results and exhibiting excellent linearity.
To achieve high-frequency inductance and impedance enhancement for effective electromagnetic interference (EMI) mitigation in power electronics, this paper presents an electromagnetic analysis and experimental study of laminated Mn-Zn toroidal ferrite cores. The electromagnetic field is analyzed using a 2D analytical solution based on a simplified Cartesian approximation. Although neglecting curvature, this approach enables efficient eigenfunction expansion and is rigorously validated against cylindrical finite difference (FDM) and 3D finite element (FEM) benchmarks. The results demonstrate that lamination effectively interrupts eddy current loops; notably, a four-layer structure increases the resonant frequency by approximately 2.8 times compared to a monolithic core. Experimental measurements confirm that this design significantly mitigates the skin effect and extends the stable frequency bandwidth. This study establishes a validated, computationally efficient methodology for optimizing core geometries to prevent impedance degradation.