We fabricated a 5 m long position-sensitive plastic scintillation optical fiber (PSOF) bundle detector composed of a sensing probe, two photomultiplier tubes (PMTs), two fast amplifiers, and a digitizer. Seven PSOFs in a bundle were used as sensing probes to estimate the gamma-ray source position, and 60Co, an uncollimated solid-disc-type radioactive isotope, was used as a gamma-ray emitter. To improve on the spatial resolution of previous studies, the transit time spread (TTS) was reduced by using a high-timing-response PMT and a bundle type of multi-cladded PSOFs. Noise was filtered out of the data. In addition, the accuracy of the data was improved through cubic spline interpolation. We determined the measurement time and measured the full width at half maximum (FWHM) considering the spatial resolution. We obtained the best spatial resolution—compared to the results of earlier studies—using our proposed bundle detector. Moreover, the sensitivity of the PSOF bundle detector was evaluated at several positions in the sensing probe. Based on the results of this study, a position-sensitive PSOF bundle detector could be used to measure gamma-ray source positions accurately over a wide contaminated area and in a shorter period of time.
In this study, we developed a remote gamma-ray spectroscopy system based on a fiber-optic radiation sensor (FORS) that is composed of an inorganic scintillator coated with reduced graphene oxide (RGO) and a plastic optical fiber (POF). As a preliminary experiment, we measured the transmitted light intensities using RGO membranes of different thicknesses with different wavelengths of emitted light. To evaluate the FORS performance, we determined the optimal thickness of the RGO membrane and measured the amounts of scintillating light and gamma energy spectra using radioactive isotopes such as 60Co and 137Cs. The amounts of scintillating light from the RGO-coated inorganic scintillators increased, and the energy resolutions of the gamma-ray spectra were enhanced. In addition, the gamma-ray energy spectra were measured using different types of RGO-coated inorganic scintillators depending on the lengths of the POFs for remote gamma-ray spectroscopy. It was expected that inorganic scintillators coated with RGO in FORS can deliver improved performance, such as increments of scintillating light and energy resolution in gamma-ray spectroscopy, and they can be used to identify nuclides remotely in various nuclear facilities.
The plastic scintillation optical fiber (PSOF) detector, characterized by its large contact area with measurement targets, effectively detects and quantifies radiation in diverse radiation-contaminated areas and liquid environments. While it is extensively utilized for measuring alpha, beta, gamma, and neutron radiations, comprehensive documentation on the spectrum measurement and energy calibration methods for gamma nuclides has not been reported. Accurate energy calibration is crucial for the precise quantification of radiation doses from various sources. The pulse-height spectrum produced by the PSOF detector does not display a Compton maximum because of the significant Gaussian energy broadening. Additionally, this spectrum compresses as the distance increases between the radiation source and the light measurement device. In this study, the energy spectrum of a PSOF for gamma nuclides was characterized by energy calibrations using Compton edge (CE). The CE channel in the measurement spectrum of the PSOF detector for three gamma nuclides was identified using the first-order differentiation method. This technique was successfully applied to spectra measured at various radiation source positions to determine the attenuation coefficients. The proposed energy calibration method allows for the conversion of pulse-height spectra obtained from alpha, beta, and neutron radiation measured with PSOF detectors into energy spectra.
Radiation possesses inherent physical characteristics, such as penetrability and radionuclide energy, which enable its widespread applicability in fields such as medicine, industry, environment, security, and research. Advancements in scintillator-based radiation detection technology have led to revolutionary changes by ensuring the safe use and precise measurement of radiation. Nevertheless, certain fields require higher scintillation yields to obtain more refined and detailed results. Therefore, in this study, we explored inorganic scintillators coated with perovskite nanomaterials to detect gamma rays with high light yields. By mixing perovskite with a polymer, we improved the intrinsic characteristics of quantum dots, which otherwise failed to maintain their performance over time. On this basis, we investigated the interactions among inorganic scintillators and a mixed material (CsPbBr3 + PMMA) and confirmed an increase in the scintillation yield and measurement trends. Furthermore, optimized scintillation yield measurement experiments facilitated gamma spectroscopy, demonstrating the validity of our approach through the analysis of the peak channel increases in the energy spectra of various gamma sources in relation to the increased scintillation yield.
In this study, a two-dimensional gamma-ray source localization system using a mesh array of plastic scintillating fibers and an artificial neural network is presented. The system covers a 200 cm by 100 cm area using SCSF-78 multi-cladded fibers. A novel U-shaped fiber topology connects both fiber ends to one side, requiring only two data-acquisition systems. Silicon photomultiplier arrays measure fast time-of-flight under optimized operating conditions to maximize signal yield. An independent artificial neural network model map measured time-of-flight values to spatial coordinates, compensating for systematic non idealities. Performance was validated using a Cesium-137 source at 20 random test positions. The artificial neural network method achieved a mean full-scale error of 4.6%. This demonstrated a 79.34% accuracy improvement over direct theoretical calculation, which had a mean full-scale error of 22.5%. The system showed consistent performance, achieving a two-dimensional standard deviation of 0.492 cm during repeatability assessment. This methodology provides a practical, efficient approach to two-dimensional radiation source localization suitable for real time monitoring and contamination mapping.