This study introduces a novel approach to analysing the combustion process using a high-speed, non-contact, optical fibre-coupled Si avalanche photodiode (APD)-based infrared radiation thermometer (IRT). The Si APD-IRT, combined with an optimised field-programmable gate array (FPGA)-based digital design, achieves a response time of 1 µs, faster than commercially available instruments. Our instrument captures the entire ignition and reignition cycle of a Jet A kerosene droplet with high temporal precision within a combustion chamber, a feat impossible with traditional thermocouples. The FPGA module was validated with a 1 µs data acquisition time, using a 40 MHz onboard clock, achieving throughput of 0.64 Gbps with efficiencies of 0.062 Mbps/slice in lookup tables (LUTs), confirming a low-area design compared to conventional FPGAs. The IRT achieves a root mean square (RMS) noise specification of 0.5 °C at a 1 µs acquisition time and a target temperature of approximately 1000 °C. A measurement uncertainty of within ±0.25% °C + 2 °C confirms that it lies within the bounds of commercial instrumentations. Our instrument was demonstrated to capture transient temperature fluctuations during combustion and characterises Jet A kerosene fuel droplets, laying the foundation for understanding sustainable aviation fuels (SAFs) and their role in transitioning from aviation fossil fuels, enabling effective research and development.
This study presents a novel approach to analysing the early stages of the combustion process by measuring the surface temperature of a kerosene droplet from its point of ignition through to its evaporation. An indium arsenide antimonide (InAsSb) photodiode-based infrared radiation thermometer (IRT), operating between 3 μm and 11 μm in wavelength, was designed to enable non-contact, low-temperature sensing with an acquisition time of 500 μs. Integrated with a data acquisition unit (DAQ), the instrument captures the transient combustion stages occurring below the droplet’s boiling point of 300 °C. The instrument was assessed against industry standards and demonstrated a measurement uncertainty of ±2 °C, confirming suitability within the performance bounds of commercial instrumentation. The IRT was deployed to measure the temperature of a kerosene droplet within an enclosed combustion chamber upon ignition, in direct comparison with a contact thermocouple. The instrument demonstrated its capability to measure the droplet’s surface temperature changes throughout its early-stage combustion. Furthermore, the wavelength specificity of the IRT eliminates thermal interference from the subsequent flame, a capability which contact thermocouples lack, thereby enabling measurement of the droplet’s temperature in isolation. This study focuses on single-droplet Jet A kerosene combustion under controlled conditions, using a transferable methodology adaptable to other fuels and environments. It supports the use of IRT for non-contact temperature measurement of fuel droplets and early-stage combustion, aiding fuel characterisation and the development of future fuels such as SAF.