Optical fiber-based acoustic detectors for ultrasound imaging in medical field feature plano-concave Fabry–Perot cavities integrated on fiber tips, realized via dip-coating. This technique imposes constraints on sensor geometry, potentially limiting performance. Lab-on-Fiber technology enables complex three-dimensional structures with precise control over geometric parameters, such as the curvature radius. A careful investigation of the optical and mechanical aspects involved in the sensors’ performances is crucial for determining the design rules of such probes. In this study, we numerically analyzed the impact of curvature on the optical and acoustic properties of a plano-concave cavity using the Finite Element Method. Performance metrics, including sensitivity, bandwidth, and directivity, were compared to planar Fabry–Perot configurations. The results suggest that introducing curvature significantly enhances sensitivity by improving light confinement, especially for cavity thicknesses exceeding half the Rayleigh zone (∼45 μm), reaching an enhancement of 2.5 a L = 60 μm compared to planar designs. The curved structure maintains high spectral quality (FOM) despite 2% fabrication perturbations. A mechanical analysis confirms no disadvantages in acoustic response and bandwidth (∼40 MHz). These findings establish curved plano-concave structures as robust and reliable for high-sensitivity polymeric lab-on-fiber ultrasound detectors, offering improved performance and fabrication tolerance for MHz-scale bandwidth applications.
The development of minimally invasive diagnostic devices in the biomedical field has grown significantly, especially those that take advantage of photoacoustic phenomena. Photoacoustic imaging is an imaging technique that exploits the photoacoustic effect, relying on the conversion of absorbed light into ultrasound waves. Thanks to lab-on-fiber technology, optical fiber can be functionalized to generate and receive a photoacoustic signal. Weak acoustic signals often limit this process, as conversion efficiency can be influenced by factors such as tissue heterogeneity, light scattering, and the attenuation of the acoustic waves within tissues. Consequently, there is significant interest in the development of highly sensitive systems with broad bandwidths. While the literature has largely focused on standard devices utilizing the interferometric effect in homogeneous slabs, this study explores the potential of multilayer structures that leverage Bragg reflection to be realized on the fiber tip. We numerically investigated both periodic and aperiodic designs of polymeric multilayer structures to further enhance the optical performance of opto-acoustic sensors. We demonstrate an enhancement in sensitivity of up to about three orders of magnitude without compromising bandwidth. This work highlights the advantages of multilayer sensor designs in improving sensitivity and performance for high-frequency opto-acoustic sensing.