This research aimed to quantitatively bioanalyze 6-mercaptopurine (6-MP), 6-methylmercaptopurine (6-MMP), and 6-thioguanosine-5′-monophosphate (6-TGMP) in dried blood spots (DBS) prepared from a small volume of acute lymphoblastic leukemia (ALL) patients. Analytes on the DBS card were extracted using 90% methanol with 5-fluorouracil (5-FU) as an internal standard. Analytical separation was performed on a Waters Acquity® UPLC BEH AMIDA column of 1.7 μm (2.1 × 100 mm) with a mobile phase mixture of 0.2% formic acid in water and 0.1% formic acid in acetonitrile-methanol, with gradient elution and a flow rate of 0.2 mL/min. Mass detection of 6-MP, 6-MMP, 6-TGMP, and 5-FU showed m/z values of 153.09 > 119.09, 167.17 > 126.03, 380.16 > 168.00, and 129.09 > 42.05, respectively. This DBS method had a run time of 5 min and yielded a linear calibration curve over a range of 25.5–1020 ng/mL for 6-MP, 6-MMP, and 6-TGMP. Analyte analysis in 22 of 24 ALL patients showed that the measured value of 6-TGMP as an active metabolite was in the range of 29–429 pmol/8 × 108 erythrocytes. Five of 22 patients had concentrations in a therapeutic range, which indicates that the treatment is effective, while 17 of 24 patients had concentrations below the therapeutic range, which indicates that a treatment dose adjustment is needed. The measured value of 6-MMP, an inactive metabolite, was in the range of 28–499 pmol/8 × 108 erythrocytes, which includes concentrations below the hepatotoxic range. The method employed here can thus be effectively utilized to support therapeutic drug monitoring.
6-Mercaptopurine is a chemotherapeutic agent of the antimetabolite class. This study aims to analyze simultaneous validation of 6-mercaptopurine (6-MP), 6-methylmercaptopurine (6-MMP), and 6-thioguanosine-5’-monophosphate (6-TGMP) in dried blood spot (DBS) using ultra performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS). An accurate volume of 60 μL blood was spotted onto DBS-CAMAG paper and then extracted using methanol 90% (v/v) containing an internal standard of 5-fluorouracil (5-FU). Separation was performed using a Waters Acquity UPLC BEH AMIDA column 1.7 μm (2.1 x 100 mm) with a mobile phase mixture of 0.2% (v/v) formic acid in water−0.1% (v/v) formic acid in acetonitrile-methanol with gradient elution and flow rate of 0.2 mL/min. Mass detection was done using Waters Xevo TQD with positive electrospray ionization (ESI) for 6-MP, 6-MMP, 6-TGMP and negative ESI for 5-FU, in multiple reaction monitoring mode. Detection rates of 6-MP, 6-MMP, 6-TGMP and 5-FU were m/z 153.09 > 119.09; 167.17 > 126.03; 380.16 > 168.00); 129.09 > 42.05, respectively. This method is linear across the range 25.5–1020 ng/mL for 6-MP, 6-MMP and 6-TGMP. This method is valid for the in vitro simultaneous analysis of 6-MP, 6-MMP and 6-TGMP in DBS, based on European Medicine Agency guidelines.