Feline infectious peritonitis (FIP), a devastating disease with near-complete mortality, is caused by the feline coronavirus (FCoV) and affects domestic cats worldwide. Herein, we report the development of a reverse transcription loop-mediated isothermal amplification (RT-LAMP) assay incorporating xylenol orange (XO) as a visual indicator for FCoV detection. The assay employed six oligonucleotide primers targeting regions of the nucleocapsid (N) gene. Under optimized conditions (65 °C, 60 min), amplification products were detected through pH-dependent colour changes in the XO dye. The RT-LAMP-XO assay exhibited high specificity for FCoV, with no cross-reactivity against other common feline viral pathogens. While the detection limit (1.7 × 101 copies/µL) was an order of magnitude higher than that of qPCR, the method offered advantages in simplicity and speed compared to existing diagnostic approaches. Although less sensitive than qPCR, the RT-LAMP-XO assay may serve as a rapid screening tool when used in combination with additional primer sets. These findings demonstrate the potential utility of XO-based RT-LAMP as a simple, visual detection method for FCoV infection.
Feline leukemia virus (FeLV) is a retrovirus that globally affects both domestic and wild cats, leading to the development of leukemia, lymphoma, and immunosuppression. However, it is important to note that the daily antigen test may yield false negative results. In this study, we successfully developed the first colorimetric loop-mediated isothermal amplification (LAMP) associated with neutral red (NR-LAMP) for the detection of FeLV. The NR-LAMP assay exhibited high sensitivity and efficiency compared to the routine polymerase chain reaction (PCR) reference method. To ensure specificity, a novel LAMP primer set was custom-designed based on the pol gene of multiple FeLV strains, which resulted in no cross-amplification with other feline viruses. Under the optimized isothermal amplification conditions at 61 °C for 40 min, the NR-LAMP assay achieved a detection limit of 100 copies/µL. Using a blind clinical test involving 98 samples, the NR-LAMP assay demonstrated perfect agreement with the reference PCR method, providing a sensitivity of 97.3% and a specificity of 100%. This proposed NR-LAMP assay surpasses other related approaches in terms of sensitivity, efficiency, and cost-effectiveness. Consequently, the colorimetric NR-LAMP reaction serves as a robust and convenient diagnostic tool for the inspection of FeLV, offering an alternative molecular method for future clinical applications and commercial utilization.