An improved method for the determination of ammonium in seawater was established using purge-and-trap pretreatment coupled with flow injection fluorescence. An appropriate volume of sodium hydroxide was added to seawater to convert ammonium to gaseous ammonia that was purged from the solution to the vapor phase by argon at 350 mL/min for 15 min at 80°C and trapped in hydrochloric acid. The solution was analyzed based on the fluorescence detection of the reaction product of o-phthaldialdehyde, sulfite, and ammonium. Parameters affecting the purge-and-trap efficiency and determination were optimized, and the matrix effects of seawater were investigated. The influence of primary amines and amino acids was tested and negligible effects were found. The detection limit was 7.4 nmol/L. The relative standard deviation was 4.4% for a seawater fortified with 200 nmol/L ammonium. Compared to the use of gas diffusion cell, this approach offers higher sensitivity, higher collection efficiency, and negligible effects from salinity. In addition, deionized water was used for standard preparation instead of matrix-matched standards.
Nitrate is the main form of dissolved inorganic nitrogen, playing an important role in both marine biogeochemical research and water environment management. In this work, the most commonly used coppered cadmium column was modified and a novel on-line Cu/Cd column with a spiral structure coupled with a de-bubbling device and syringe-type filter was developed. With the advantages of convenience, portability, stability, and high reduction efficiency, the interference of air bubbles in the column could easily be avoided. Based on the classic Griess reaction, a simple reverse flow injection system coupled with a novel Cu/Cd column and custom-made flow cell was established for the field spectrophotometric determination of nitrate in seawater. The effects of certain reaction parameters—including the reagent concentration, flow rate, length of the Cu/Cd column, and salinity—were investigated, optimized, and compared with pure water, with an approximate 9% increase in the sensitivity of seawater samples. This method exhibited a detection limit of 0.03 μmol/L, with a relative standard deviation of 0.6%, and the working range was 20 μmol/L before dilution. Compared to the referred vanadium reduction method based on the same flow system, the proposed method showed significant advantages including sensitivity and reproducibility. No significant difference was observed between the analytical seawater sample results obtained by the proposed and reference methods. Furthermore, the proposed method was validated by the first class of National Certified Reference Materials and successfully applied to the nitrate determination of Wenling coastal seawater (Zhejiang, China).