The accurate monitoring of flow velocity is crucial in applications such as blood microcirculation and microfluidic systems. However, the high sensitivity of current hot wire flowmeters is often achieved at the expense of increasing the initial temperature, which imposes significant limitations when measuring blood or other temperature sensitive fluids. In this study, a fiber sensor probe with a plano-concave cavity, fabricated from a PbS quantum dots (QDs)-doped photoresist, is proposed for the sensitive flow velocity detection of microfluidics. In the proposed hot wire-based micro-flowmeter, the excitation laser (980 nm) is efficiently absorbed and converted into thermal energy, while minimally affecting the high-quality interference of the cavity at the C-band. The experimental results show that only a 3 °C increase in temperature is required for flow velocity monitoring, with a sensitivity of 7.7 pm/(mm/s) achieved within a linear response range of 3.82 mm/s to 16.72 mm/s. Additionally, an intensity interrogation scheme is introduced for the hot wire-based fiber sensor probe. This low initial temperature requirement makes the proposed sensor suitable for microfluidics, demonstrating promising potential for use in microcirculation measurement and drug delivery systems.
Optic-fiber-based thermal flowmeters have the merits of compact size and high sensitivity, which typically require two light beams separately acting as a pump for heating the sensing unit and a probe for sensing temperature with the variation of external flow. Here, we propose a metallic nanostructure with multiple plasmonic resonance modes for the application of an optic-fiber-based thermal flowmeter. The optical properties of a nanostructure comprised of a double-width gold grating, a poly (methylmethacrylate) (PMMA) layer, and a gold film are numerically simulated in the spectral range of 600–1800 nm. The optical resonances of different modes are systematically investigated with the variation of the structural parameters. Interestingly, two optical resonance modes with distinct spectral shift under the same temperature variation, i.e., 21.34 pm/°C vs. 269.2 pm/°C, are obtained after the strategic optimization of the nanostructure. Finally, the sensitivity of the flowmeter with the proposed nanostructure is investigated by adopting the low-temperature sensitivity mode for optical pumping and the high-temperature sensitivity mode for temperature sensing, proving its significant potential as an optic-fiber-based thermal flowmeter.