Background: Multielectrode mapping catheters, such as the OCTARAY and OPTRELL, are essential in creating myocardial electroanatomical mapping in arrhythmias. The OCTARAY is a multi-spline mapping catheter with 48 closely spaced multielectrodes that enables high-resolution electroanatomical mapping, while the OPTRELL is a multi-electrode catheter with 36 electrodes arranged on eight radiating splines. However, only a few studies have compared their performance. In this study, we aimed to compare the OCTARAY and OPTRELL catheters in two areas: left atrial (LA) mapping during atrial fibrillation (AF) ablation; and premature ventricular contraction (PVC) mapping. Methods: Twenty patients (Ten patients for LA mapping and ten for PVC mapping) were enrolled. LA voltage mapping was performed twice, alternating between catheters post-AF ablation. Parameters compared included mapping time, mapping points, catheter-induced premature atrial contraction (PACs), tissue proximity indication, low voltage area, and fluoroscopy time. For PVC mapping, comparisons included mapping time, catheter-induced PVCs, earliest activation time measured from the onset of PVC QRS, earliest activation point, and fluoroscopy time. Results: Compared with THE OCTARAY, mean voltage using the OPTRELL was higher (0.192 mV[0.072, 0.48] vs. 0.126 mV[0.042, 0.378]; P = .001) and the percentage of tissue proximity indication positive was also higher (14.97% vs. 11.45%; P < .001). However, there were no significant differences in low voltage area between the two groups (39.5 m 2 [16.5, 66.8] vs. 40[23.6, 61]; P = .861), and in other LA parameters. In PVC mapping, catheter-induced PVCs using OPTRELL were significantly fewer than the OCTARAY (100 [32, 337] vs. 247 [110, 745], P = .039), with fewer catheter induced PVCs per minute (15 [6, 23] vs. 35 [20, 71], P = .039). However, no significant differences were observed in other PVCs mapping parameters. Conclusion: The OPTRELL catheter demonstrated higher voltage recordings in LA mapping and fewer catheter-induced PVCs compared with the OCTARAY catheter. However, no significant difference was observed in other mapping parameters.