This study aimed to determine the intra- and inter-device accuracy and reliability of wearable athletic tracking devices, under controlled laboratory conditions. A total of nineteen portable accelerometers (Catapult OptimEye S5) were mounted to an aluminum bracket, bolted directly to an Unholtz Dickie 20K electrodynamic shaker table, and subjected to a series of oscillations in each of three orthogonal directions (front-back, side to side, and up-down), at four levels of peak acceleration (0.1g, 0.5g, 1.0g, and 3.0g), each repeated five times resulting in a total of 60 tests per unit, for a total of 1140 records. Data from each accelerometer was recorded at a sampling frequency of 100Hz. Peak accelerations recorded by the devices, Catapult PlayerLoad™, and calculated player load (using Catapult’s Cartesian formula) were used for the analysis. The devices demonstrated excellent intradevice reliability and mixed interdevice reliability. Differences were found between devices for mean peak accelerations and PlayerLoad™ for each direction and level of acceleration. Interdevice effect sizes ranged from a mean of 0.54 (95% CI: 0.34–0.74) (small) to 1.20 (95% CI: 1.08–1.30) (large) and ICCs ranged from 0.77 (95% CI: 0.62–0.89) (very large) to 1.0 (95% CI: 0.99–1.0) (nearly perfect) depending upon the magnitude and direction of the applied motion. When compared to the player load determined using the Cartesian formula, the Catapult reported PlayerLoad™ was consistently lower by approximately 15%. These results emphasize the need for industry wide standards in reporting validity, reliability and the magnitude of measurement errors. It is recommended that device reliability and accuracy are periodically quantified.
Skeletal fractures associated with bone mass loss are a major clinical problem and economic burden, and lead to significant morbidity and mortality in the ageing population. Clinical image-based measures of bone mass show only moderate correlative strength with bone strength. However, engineering models derived from clinical image data predict bone strength with significantly greater accuracy. Currently, image-based finite element (FE) models are time consuming to construct and are non-parametric. The goal of this study was to develop a parametric proximal femur FE model based on a statistical shape and density model (SSDM) derived from clinical image data. A small number of independent SSDM parameters described the shape and bone density distribution of a set of cadaver femurs and captured the variability affecting proximal femur FE strength predictions. Finally, a three-dimensional FE model of an ‘unknown’ femur was reconstructed from the SSDM with an average spatial error of 0.016 mm and an average bone density error of 0.037 g/cm3.