Several measurements of the radon concentration are performed by RAD7 in the University of South China. We find that 30–40 min is needed for RAD7 for tracing the concentration of the standard radon chamber. There are two reasons. The first is that the sufficient time of air cycle is needed for the radon concentration in internal cell of RAD7 equal to that of the environment; and the second is that the sufficient decay time is needed for the 218Po concentration in internal cell of RAD7 equal to that of the radon. We used a zeroth order approximation to describe the evolution of the environment radon concentration, and obtained a novel algorithm for quick and continuous tracing the change of radon concentration. The corrected radon concentration obtained through this method is in good agreement with the reference value. This method can be applied to develop and improve the instruments for tracing the change of radon concentration quickly.
Radon-222 (Rn-222) exhalation rate is vital for estimating radiation risk from many kinds of materials. AlphaGUARD measures the radon concentration based on the ionization chamber principle, which is currently recognized as a reference instrument to measure radon. In China, measurements of radon exhalation rate are performed by AlphaGUARD operated in flow-through mode on a reference device to verify measurement accuracy. These measurements are performed in both open and closed loop. AlphaGUARD can fast rapidly the variation of the radon concentration in the chamber, which is tightly pressed against the surface of the medium to accumulate the exhaled radon. When the model is used to obtain the radon exhalation rate, the radon exhalation rates obtained by nonlinear data fitting on the measured radon concentrations are similar to the reference value of the device. The difference of radon exhalation rate values of six measurements is small.
We propose a novel method for quickly measuring the radon exhalation rate in open loop. We first obtain the temporal variation of radon concentration in the internal cell of the RAD7 by analyzing the work principle of RAD7. We then obtain the temporal variation of radon concentration in the ventilation-type accumulation chamber when the effects of leakage and back diffusion are neglected. This method uses the measured value before the radon concentration in the ventilation-type accumulation chamber reaches a steady state. The diameter of the air input tube to the ventilation-type accumulation is large enough to keep the differential pressure in the accumulation chamber and outdoors negligible. Short cycle time and large flow rate will be appropriate for reducing measurement error. Several radon exhalation rate measurements of the medium surface have been performed in the Radon Laboratory of the University of South China. The radon exhalation rates obtained by verification experiments are in good agreement with the reference value. This method can be applied to develop and improve the instruments for measuring radon exhalation rate.
A novel method was proposed to measure the radon exhalation rate in only one measurement cycle. We obtain the radon concentration variety regulation in the internal cell of the RAD7 by analyzing the work principle of RAD7 and the radon concentration variety regulation in the ventilation-type accumulation chamber when the effects of leakage and back diffusion are neglected. This method uses the measured value before the radon concentration in the ventilation-type accumulation chamber reaches a steady state. Several radon exhalation rate measurements of the medium surface have been performed in the Radon Laboratory of the University of South China. The radon exhalation rates obtained by verification experiments are in good agreement with the reference value. This simple, accurate and fast method can be applied to develop and improve the instruments for measuring radon exhalation rates.