On the basis of positive clinical results with mid- and long-term follow-up using the Mayo short stem, the Metha neck-preserving stem (BBraun, Aesculap, Tuttlingen, Germany) was introduced. The purpose of this study was to validate the implant design by direct acquisition of bone remodelling data from total hip arthroplasty (THA) recipients using dual-energy X-ray absorptiometry (DEXA). After power analysis, 25 patients were included in this prospective study. Patients were examined clinically and underwent DEXA examinations preoperatively and postoperatively at one week, six months and one and two years after THA. Gruen zones were adapted to the short stem design (R1–R7). The Harris Hip Score (HHS) increased significantly by 31 points. No stem had to be revised. Bone mineral density (BMD) in the greater trochanter decreased significantly from 0.78 g/cm2 postoperatively to 0.72 g/cm2 two years after surgery. Marginal changes were seen in the lateral distal regions (R4–R5). In the minor trochanter region, BMD increased significantly after two years by 12.9%. In the calcar region, BMD exceeded the baseline value by 6.1% two years after implantation. Stress shielding seems to occur at the greater trochanter due to the vast cross-section of the implant. However, the aim of proximal load transfer of the Metha stem seems to be partially achieved. DEXA analysis revealed a concentrated load distribution on the medial portion of the femur, which is an important region to guarantee long-term implant survival.
Background: There are several numerical investigations on bone remodelling after total hip arthroplasty (THA) on the basis of the finite element analysis (FEA). For such computations certain boundary conditions have to be defined. The authors chose a maximum of three static load situations, usually taken from the gait cycle because this is the most frequent dynamic activity of a patient after THA.Materials and methods: The numerical study presented here investigates whether it is useful to consider only one static load situation of the gait cycle in the FE calculation of the bone remodelling. For this purpose, 5 different loading cases were examined in order to determine their influence on the change in the physiological load distribution within the femur and on the resulting strain-adaptive bone remodelling. First, four different static loading cases at 25%, 45%, 65% and 85% of the gait cycle, respectively, and then the whole gait cycle in a loading regime were examined in order to regard all the different loadings of the cycle in the simulation.Results: The computed evolution of the apparent bone density (ABD) and the calculated mass losses in the periprosthetic femur show that the simulation results are highly dependent on the chosen boundary conditions.Conclusion: These numerical investigations prove that a static load situation is insufficient for representing the whole gait cycle. This causes severe deviations in the FE calculation of the bone remodelling. However, accompanying clinical examinations are necessary to calibrate the bone adaptation law and thus to validate the FE calculations.