Cortisol is commonly used as a significant biomarker of psychological or physical stress. With the accelerated pace of life, non-invasive cortisol detection at the point of care (POC) is in high demand for personal health monitoring. In this paper, an ultrasensitive immunosensor using gold nanoparticles/molybdenum disulfide/gold nanoparticles (AuNPs/MoS2/AuNPs) as transducer was explored for non-invasive salivary cortisol monitoring at POC with the miniaturized differential pulse voltammetry (DPV) system based on a smartphone. Covalent binding of cortisol antibody (CORT-Ab) onto the AuNPs/MoS2/AuNPs transducer was achieved through the self-assembled monolayer of specially designed polyethylene glycol (PEG, SH-PEG-COOH). Non-specific binding was avoided by passivating the surface with ethanolamine. The miniaturized portable DPV system was utilized for human salivary cortisol detection. A series current response of different cortisol concentrations decreased and exhibited a linear range of 0.5–200 nM, the detection limit of 0.11 nM, and high sensitivity of 30 μA M−1 with a regression coefficient of 0.9947. Cortisol was also distinguished successfully from the other substances in saliva. The recovery ratio of spiked human salivary cortisol and the variation of salivary cortisol level during one day indicated the practicability of the immunosensor based on the portable system. The results demonstrated the excellent performance of the smartphone-based immunosensor system and its great potential application for non-invasive human salivary cortisol detection at POC.
Personalized, point-of-care testing of human tears is essential for ocular disease diagnosis, yet it is hampered by picomolar biomarker levels and microliter sample volumes. In this work, we developed an integrated, portable electrochemiluminescence (ECL)-based device for rapid and quantitative detection of tumor necrosis factor alpha (TNF-α), a pivotal inflammatory marker in ocular surface disease, with particular relevance to dry eye syndrome (DES). The device integrates a miniaturized electrochemical cell for ECL reactions and a compact silica photomultiplier for signal measurement. A vertical silica mesochannel (VSM)-coated ITO electrode is also integrated and further functionalized with TNF-α-specific aptamers. The VSM enables the enrichment of ECL luminophores, thus enabling further amplification of ECL signals and enhancing sensitivity. A wide linear range from 0.1 to 200 pg/mL was achieved using 10-fold dilution of 3 μL tear samples. Overall, this study provides a portable, highly sensitive platform for personalized analysis of TNF-α in tear fluid, enabling rapid point-of-care assessment of DES.