Xiaobing Pang

Dr. Xiaobing Pang

College of Environment, Zhejiang University of Technology, 18 Chaowang Road, Hangzhou, China.

Research Focus & Metrics

Academic Career Length

16Years (2011-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
1 November 2011 in Science of The Total Environment
Verified
Journal Article
Closed access
XPXiaobing Pang
Xiaobing Pang
ALAlastair C. Lewis
Alastair C. Lewis

1 April 2021 in Science of The Total Environment
Verified
Journal Article
Closed access
XPXiaobing Pang
Xiaobing Pang
LCLang Chen
Lang Chen
KSKangli Shi
Kangli Shi
FWFei Wu
Fei Wu
JCJianmeng Chen
Jianmeng Chen
MXMeng Xu
Meng Xu

1 July 2011 in Talanta
Verified
Journal Article
Closed access
XPXiaobing Pang
Xiaobing Pang
ALAlastair C. Lewis
Alastair C. Lewis
JHJacqueline F. Hamilton
Jacqueline F. Hamilton

1 June 2013 in Journal of Chromatography A
Verified
Journal Article
Closed access
XPXiaobing Pang
Xiaobing Pang
ALAlastair C. Lewis
Alastair C. Lewis
MGMILA RÓDENAS GARCÍA
MILA RÓDENAS GARCÍA

3 February 2014 in Atmospheric Measurement Techniques
Verified
Journal Article
Open access
PDF file
XPXiaobing Pang
Xiaobing Pang
ALA. C. Lewis
A. C. Lewis
ARA. R. Rickard
A. R. Rickard
MBM. T. Baeza-Romero
M. T. Baeza-Romero
TAT. J. Adams
T. J. Adams
AMAmalia Muñoz
Amalia Muñoz
A microfluidic lab-on-a-chip derivatisation technique has been developed to measure part per billion (ppbV) mixing ratios of gaseous glyoxal (GLY) and methylglyoxal (MGLY), and the method is compared with other techniques in a smog chamber experiment. The method uses o-(2, 3, 4, 5, 6-pentafluorobenzyl) hydroxylamine (PFBHA) as a derivatisation reagent and a microfabricated planar glass micro-reactor comprising an inlet, gas and fluid splitting and combining channels, mixing junctions, and a heated capillary reaction microchannel. The enhanced phase contact area-to-volume ratio and the high heat transfer rate in the micro-reactor resulted in a fast and highly efficient derivatisation reaction, generating an effluent stream ready for direct introduction to a gas chromatograph-mass spectrometer (GC-MS). A linear response for GLY was observed over a calibration range 0.7 to 400 ppbV, and for MGLY of 1.2 to 300 ppbV, when derivatised under optimal reaction conditions. The analytical performance shows good accuracy (6.6% for GLY and 7.5% for MGLY), suitable precision (<12.0%) with method detection limits (MDLs) of 75 pptV for GLY and 185 pptV for MGLY, with a time resolution of 30 min. These MDLs are below or close to typical concentrations of these compounds observed in ambient air. The feasibility of the technique was assessed by applying the methodology to quantify α-dicarbonyls formed during the photo-oxidation of isoprene in the EUPHORE chamber. Good correlations were found between microfluidic measurements and Fourier Transform InfraRed spectroscopy (FTIR) with a correlation coefficient (r2) of 0.84, Broadband Cavity Enhanced Absorption Spectroscopy (BBCEAS) (r2 = 0.75), solid phase micro extraction (SPME) (r2 = 0.89), and a photochemical chamber box modelling calculation (r2 = 0.79) for GLY measurements. For MGLY measurements, the microfluidic technique showed good agreement with BBCEAS (r2 = 0.87), SPME (r2 = 0.76), and the modeling simulation (r2 = 0.83), FTIR (r2 = 0.72) but displayed a discrepancy with Proton-Transfer Reaction Time-of-Flight Mass Spectrometry (PTR-ToF-MS) with r2 value of 0.39.

29 December 2016 in Journal of Separation Science
Verified
Journal Article
Open access
PDF file
XPXiaobing Pang
Xiaobing Pang
ALAlastair C. Lewis
Alastair C. Lewis
MSMarvin D. Shaw
Marvin D. Shaw
Stir bar sorptive extraction is a powerful technique for the extraction and analysis of organic compounds in aqueous matrices. Carbonyl compounds are ubiquitous components in rainwater, however, it is a major challenge to accurately identify and sensitively quantify carbonyls from rainwater due to the complex matrix. A stir bar sorptive extraction technique was developed to efficiently extract carbonyls from aqueous samples following chemical derivatization by O‐(2,3,4,5,6‐pentafluorobenzyl) hydroxylamine hydrochloride. Several commercial stir bars in two sizes were used to simultaneously measure 29 carbonyls in aqueous samples with detection by gas chromatography with mass spectrometry. A 100 mL aqueous sample was extracted by stir bars and the analytes on stir bars were desorbed into a 2 mL solvent solution in an ultrasonic bath. The preconcentration Coefficient for different carbonyls varied between 30 and 45 times. The limits of detection of stir bar sorptive extraction with gas chromatography mass spectrometry for carbonyls (10–30 ng/L) were improved by ten times compared with other methods such as gas chromatography with electron capture detection and stir bar sorptive extraction with high‐performance liquid chromatography and mass spectrometry. The technique was used to determine carbonyls in rainwater samples collected in York, UK, and 20 carbonyl species were quantified including glyoxal, methylglyoxal, isobutenal, 2‐hydroxy ethanal.