Amanda Wellman

Amanda Wellman

orcid

Research Focus & Metrics

Academic Career Length

2Years (2025-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
1 May 2025 in bioRxiv
Verified
Preprint
Open access
PDF file
BOBana H. Odeh
Bana H. Odeh
AWAmanda Wellman
Amanda Wellman
MAMichael Ameye
Michael Ameye
ZAZachary Atwood
Zachary Atwood
LGLuke Gray
Luke Gray
JRJoseph A Roche
Joseph A Roche
Objective A commercial electrode used for transcutaneous neuromuscular electrical stimulation (NMES) in mice is no longer available. In response, we developed two low- cost, customizable alternatives—a 3D-Printed Electrode and a Pen Electrode assembled from off-the-shelf jumper wire components—and evaluated their performance relative to the discontinued commercial standard. Methods We conducted in vivo NMES of the left hindlimb ankle dorsiflexors in C57BL/6J mice using three electrode types: a Simple Electrode (previously available as BS4 50–6824, Harvard Apparatus), our novel 3D-Printed Electrode, and our novel Pen Electrode. Contractile torque was recorded during twitch and tetanic contractions, and electrode performance was evaluated based on peak torque values. A repeated measures ANOVA was used to compare torque generated by each animal for the three different types of electrodes. Results Both custom electrodes generated twitch and tetanic responses comparable to those of the Commercial electrode. No statistically significant differences were observed in peak torque values in either twitch (p = 0.181) or tetany (p = 0.438). Conclusion Custom-built 3D-printed and Pen-style electrodes offer low-cost, accessible alternatives to discontinued commercial devices for murine NMES studies. These tools can facilitate continued use of electrical stimulation protocols in preclinical neuromuscular research.

30 April 2026 in Muscles
Verified
Journal Article
Open access
PDF file
BOBana H. Odeh
Bana H. Odeh
AWAmanda Wellman
Amanda Wellman
MAMichael Ameye
Michael Ameye
ZAZachary Atwood
Zachary Atwood
LGLuke Gray
Luke Gray
JRJoseph A Roche
Joseph A Roche
Neuromuscular electrical stimulation (NMES) is integral to studying muscle function in healthy and dystrophic mice. Certain commercial electrodes and laboratory stimulators used for NMES in mice are no longer in production. We developed and/or tested low-cost, open-source alternatives to discontinued commercial standards. We performed two studies—a comparison of electrodes and a comparison of stimulators. In the electrode study, in vivo NMES was applied to the left hindlimb ankle dorsiflexors in healthy C57BL/6J and dysferlin-null BLAJ mice using three electrode types: a previously available commercial electrode, a custom 3D-Printed electrode, and a custom Pen electrode assembled from off-the-shelf components. Twitch and tetanic torque were measured and compared using two-way repeated-measures ANOVA. Twitch torque differed by electrode type (p = 0.031), with lower values observed for the Pen electrode compared with the 3D-Printed electrode (e.g., 573 ± 72 vs. 666 ± 70 mN.mm in C57BL/6J mice), whereas tetanic torque did not differ significantly between electrode types (p = 0.060). In the stimulator study, twitch and tetanic contractions were elicited using the open-source StimJim stimulator and compared with contractions elicited by the discontinued Grass S48 stimulator. Twitch torque was lower with the StimJim (588 ± 107 mN.mm) compared with the Grass S48 (698 ± 116 mN.mm; p < 0.001), whereas tetanic torque values were not statistically different (p = 0.055). These findings indicate that open-source electrodes and stimulators can produce similar maximal tetanic torque under the tested conditions, although differences in twitch torque and stimulation parameters should be considered. These results reflect a methodological validation of accessible tools rather than a formal equivalence analysis.