In synchrotron light source applications, the photon beam interaction with various materials can produce electric charge which can be measured as current and be used to diagnose particle trajectories, beam intensity, beam profile, position, and stability. Sirius, a 4th-generation Brazilian synchrotron light source, will make use of hundreds of low-intensity measurement instruments. This work aims to show and discuss the design details, challenges, and test results of an Ethernet four-channel high-performance digital ammeter, applied for general-purpose beam diagnostics. The device is based on low-noise and extremely low input bias bipolar transimpedance amplifiers with eight selectable ranges (full scales from pA to mA). Characterization results show that the achieved gain, accuracy, and noise performance are on the same order of magnitude as those of expensive commercial benchtop equipment. In low bandwidth applications, the device was able to measure amplitudes of hundreds of picoampere with intrinsic noise of units of femtoampere (RMS). The designed device is a cost-effective solution.
To address the high demand for precise low-current measurements at the Sirius’ beamlines, a quad-channel high-resolution Ethernet picoammeter has been designed. The instrument can measure currents ranging from femtoampere to milliampere across eight selectable ranges, featuring integrated analog-to-digital converters (ADCs), enabling sampling rates of up to 2 ksps and synchronization capabilities. This work describes the hardware design, the hardware handling procedures to achieve sub-picoampere resolution, and the characterization results of the instrument, considering the experimental results from Sirius beamlines. The designed device provides noise performance and gain accuracy that is comparable to high-end commercial solutions, proving its suitability for critical applications like on-the-fly scanning experiments. Special attention will be given to evaluating trigger latency, synchronization outcomes, as well as the device’s installation and commissioning at beamlines. Furthermore, we will deeply explore the interplay between the trigger period, digital filter bandwidth, and front-end analog bandwidth to optimize the signal-to-noise ratio in specific applications. The hardware project is publicly available in CERN’s open hardware repository.
TUPI (Timepix4-based Ultra-fast Photon Imaging) is the new family of direct conversion and vacuum-compatible hybrid detectors under design at Sirius, the Brazilian 4th generation light source. Based on the Timepix4 Application Specific Integrated Circuit (ASIC), it targets different kinds of experiments, such as tomography and spectrometry, and will be used in multiple bioimaging beamlines of Orion, Brazil's Biosafety Level 4 (BSL4) Laboratory project. Also, it will cover a wide energy range (5 to 45 keV) by using different sensors materials (Si, GaAs, CdTe, etc.). Designed in a modular approach to ensure scalability for constructing larger active areas in a compact form, each TUPI module will be composed by a set of 3x side-by-side Timepix4 ASICs to provide a sensitive area of approximately 74×28 mm2 (1344×512 pixels of 55 μm pitch), streaming up to 44 kfps through up to 48×10.24 Gbps optical links directly connected to a Data Acquisition (DAQ) system.