Wioletta Ślusarczyk-Kacprzyk

Mrs. Wioletta Ślusarczyk-Kacprzyk

orcid

Research Focus & Metrics

Academic Career Length

17Years (2010-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
17 May 2010 in Strahlentherapie und Onkologie
Verified
Journal Article
Closed access
PDF file
WSWitold Skrzyński
Witold Skrzyński
SZSylwia Zielińska-Dąbrowska
Sylwia Zielińska-Dąbrowska
MWMarta Wachowicz
Marta Wachowicz
WŚWioletta Ślusarczyk-Kacprzyk
Wioletta Ślusarczyk-Kacprzyk
PKPaweł F. Kukołowicz
Paweł F. Kukołowicz
WBWojciech Bulski
Wojciech Bulski

2 February 2018 in Polish Journal of Radiology
Verified
Journal Article
Open access
PDF file
EFEwa Fabiszewska
Ewa Fabiszewska
KPKatarzyna Pasicz
Katarzyna Pasicz
IGIwona Grabska
Iwona Grabska
WSWitold Skrzyński
Witold Skrzyński
WŚWioletta Ślusarczyk-Kacprzyk
Wioletta Ślusarczyk-Kacprzyk
WBWojciech Bulski
Wojciech Bulski
Background: Regular quality control is required in Poland only for those methods of medical imaging which involve the use of ionizing radiation but not for ultrasonography. It is known that the quality of ultrasound images may be affected by the wearing down or malfunctioning of equipment. Material/Methods: An evaluation of image quality was carried out for 22 ultrasound scanners equipped with 46 transducers. The CIRS Phantom model 040GSE was used. A set of tests was established which could be carried out with the phantom, including: depth of penetration, dead zone, distance measurement accuracy, resolution, uniformity, and visibility of structures. Results: While the dead zone was 0 mm for 89% of transducers, it was 3 mm for the oldest transducer. The distances measured agreed with the actual distances by 1 mm or less in most cases, with the largest difference of 2.6 mm. The resolution in the axial direction for linear transducers did not exceed 1 mm, but it reached even 5 mm for some of the convex and sector transducers, especially at higher depths and in the lateral direction. For 29% of transducers, some distortions of anechoic structures were observed. Artifacts were detected for several transducers. Conclusions: The results will serve as a baseline for future testing. Several cases of suboptimal image quality were identified along with differences in performance between similar transducers. The results could be used to decide on the applicability of a given scanner or transducer for a particular kind of examination.

18 August 2016 in Polish Journal of Radiology
Verified
Journal Article
Open access
PDF file
WŚWioletta Ślusarczyk-Kacprzyk
Wioletta Ślusarczyk-Kacprzyk
WSWitold Skrzyński
Witold Skrzyński
EFEwa Fabiszewska
Ewa Fabiszewska
BACKGROUND:Different methods of image quality evaluation are routinely used for analogue and digital mammography systems in Poland. In the present study, image quality for several screen-film (SF), computed radiography (CR), and fully digital (DR) ma...

8 July 2016 in Journal of Applied Clinical Medical Physics
Verified
Journal Article
Open access
PDF file
WSWitold Skrzyński
Witold Skrzyński
WŚWioletta Ślusarczyk-Kacprzyk
Wioletta Ślusarczyk-Kacprzyk
The geometric accuracy of CT images is essential when they are used for planning radiation therapy. In this study, a method of quantitative testing of image rotation in CT is presented, based on the automated analysis of an image of a phantom half-filled with water. A plug-in was written for ImageJ software for automated detection of a water surface in an image and measurement of its angle relative to a horizontal line. A water phantom can be used for the evaluation of image rotation for axial mode. In helical mode the movement of the table would cause movement of the water. The difference between image rotation for axial and helical scans can be evaluated by measuring and comparing the angles of the tabletop surface for both modes. Preliminary results of measurements for three CT scanners show that image rotation does not exceed 0.5°, and is less than 0.1° for the dedicated CT simulator. It was observed that, for one CT scanner image, the rotation angle depended on tube rotation time. PACS number(s): 87.57.Q-, 87.55.Qr, 06.30.Bp

1 November 2017 in Radiation Measurements
Verified
Journal Article
Closed access
JKJ. Kunst
J. Kunst
RKRenata Kopec
Renata Kopec
PKP. Kukołowicz
P. Kukołowicz
NMN. Mojżeszek
N. Mojżeszek
BSB. Sadowski
B. Sadowski
POP. Olko
P. Olko

18 November 2023 in Physica Medica
Verified
Journal Article
Closed access
PWPaulina Wesolowska
Paulina Wesolowska
WŚWioletta Ślusarczyk-Kacprzyk
Wioletta Ślusarczyk-Kacprzyk
MFMarta Fillmann
Marta Fillmann
PKPavel Kazantsev
Pavel Kazantsev
WBWojciech Bulski
Wojciech Bulski