Background: Extended-wear infusion sets (E-iS) allowing up to 7 days of catheter use have recently become available. Real-world evidence of contemporary automated insulin delivery (AID) systems using E-iS remains limited. Method: A prospective non-randomized single-arm real-world study in adults with type 1 diabetes using the MiniMed-780G AID system who transitioned from conventional 3-day infusion sets to E-iS was conducted. Clinical variables, device settings, adverse events, and patient-reported outcomes were assessed at baseline, 3 and 6 months. Results: 65 participants were included, 58 completed 6 months of follow-up.Mean catheter wear time increased from 3.6 ± 1.1 to 6.1 ± 2.3 days,(p < 0.001), and reservoir change interval increased from 3.5 ± 1.0 to 5.2 ± 2.3 days,(p < 0.001). Glycemic control remained overall stable, although small changes were observed in some CGM metrics (Time in range-TIR: 80.0 ± 7.1% to 78.0 ± 10.0%, p = 0.014) and slight increases in time above range, while hypoglycemia exposure remained unchanged. HbA1c decreased from 6.8 ± 0.5% to 6.4 ± 1.0%, (p = 0.03). Total daily insulin dose did not change, but insulin use related to infusion set management decreased significantly. Pain at cannula insertion improved from 2.1 ± 1.5 to 1.2 ± 1.1,(p < 0.001), and treatment satisfaction increased from 8.2 ± 1.9 to 9.0 ± 1.2,(p = 0.012). No episodes of severe hypoglycemia or diabetic ketoacidosis occurred. Cannula wear duration was strongly correlated with reservoir change interval (r = 0.712, p < 0.001). At 6 months, 92.3% of participants continued using the E-iS, with 48.3% maintaining the full 7-day wear duration. Conclusions: In real-world use with an AID system, the E-iS significantly prolonged catheter wear duration and reservoir change intervals while remaining safe and well tolerated. It was associated with high satisfaction and stable glycemic outcomes, supporting E-iS as a practical strategy to reduce the burden of infusion set changes in routine clinical practice.
Background/Objectives: Infusion sets are essential for continuous subcutaneous insulin infusion. Conventional sets require replacement every 2–3 days, while recent research has focused on extending wear time to reduce treatment burden and support automated insulin delivery. We aimed to characterize evidence on extended-wear infusion sets (>3 and ≤7 days) in people with type 1 diabetes (T1D), focusing on survival, safety, insulin delivery, glycemic outcomes, pharmacokinetic/pharmacodynamic parameters, and user experience. Methods: A scoping review was conducted according to PRISMA-ScR. The protocol was registered in the Open Science Framework (OSF; DOI: 10.17605/OSF.IO/TNKJ5). MEDLINE/PubMed, Embase, Scopus, Web of Science Core Collection, and Cochrane CENTRAL were searched for studies published from 1 January 2020 to 4 September 2026. Eligible studies evaluated infusion sets worn for >3 days in people with T1D and reported extended-wear outcomes. Results: Of 229 records identified, 103 unique references were screened and 11 underwent full-text assessment. Ten publications met the inclusion criteria, representing 9 independent sources of evidence. In the largest studies, 7-day set survival ranged from 77.8% to 95%, depending on device and failure criteria. Some studies reported deterioration in glycemic or pharmacodynamic measures with prolonged wear, whereas others showed good tolerability and reduced burden from set changes. Conclusions: Extended-wear infusion sets can remain functional for 4–7 days in a substantial proportion of people with T1D, although performance may decline with increasing wear time. Evidence is limited by heterogeneity in devices and study designs, small samples, few direct comparisons, and difficulty isolating the infusion set effect from other technological components. Independent comparative studies are needed to establish safety, insulin delivery stability, and clinical benefit.