A mathematical model is implemented to study the performance of an insulator-based dielectrophoretic device. The geometry of the device was captured in a computational model that solves Laplace equation within an array of cylindrical insulating structures. From the mathematical model it was possible to predict the location and magnitude of the zones of dielectrophoretic trapping of microparticles. Simulation and experimental results of trapping zones are compared for different operating conditions.
Salmonella is one of the main microorganisms that causes food-borne illnesses worldwide, although there are standard procedures to determine its presence or absence in food samples. However, these methodologies are time consuming and cumbersome, so the development of devices that rapidly and accurately determine bacteria is necessary. Various materials have been used to construct these devices. Hydrogenated amorphous silicon carbide (a-SiC:H) is one of the least studied. Therefore, in the present work, self-assembled monolayers were applied on a-SiC:H thin films to determine Salmonella Typhimurium ATCC 14028 by Fourier transform infrared spectroscopy. The detection of this microorganism was performed between 1060 and 960 cm−1 corresponding to carbohydrate and phosphate groups of the bacteria. The lowest detected concentration was 10 CFU/mL. This work confirms the utility of a-SiC:H to reliably and rapidly determine Salmonella.
This work presents the design of a biosensor for detection of Escherichia coli bacteria. The biosensor consists of an interdigitated electrodes array to detect the impedance changes, as a function of frequency, when bacteria is captured in different areas. The design was first simulated with COMSOL Multiphysics software to obtain the impedance magnitude and phase curves, after that the transfer function for every case was established with an algorithm in MATLAB software to propose an equivalent circuit in order to determine the electrical components that dominate the impedance response in every frequency range of the curve. Results show that impedance changes in the biosensor are mainly due to the bacteria captured between interdigitated electrodes, whereas the bacteria captured only over interdigitated electrodes produce a very small change in the impedance. In order to confine the bacteria between electrodes, a microchannel is added, increasing the biosensor sensitivity.