Deep brain stimulation (DBS) to the superolateral branch of the medial forebrain bundle (MFB) has been reported to lead to rapid antidepressant effects. In this longitudinal study, we expand upon the initial results we reported at 26 weeks (Fenoy et al., 2016), showing sustained antidepressant effects of MFB DBS on six patients with treatment-resistant depression (TRD) over 1 year. The Montgomery-Åsberg Depression Rating Scale (MADRS) was used as the primary assessment tool. Deterministic fiber tracking was used to individually map the target area; analysis was performed to compare modulated fiber tracts between patients. Intraoperatively, upon stimulation at target, responders reported immediate increases in energy and motivation. An insertional effect was seen during the 4-week sham stimulation phase from baseline (28% mean MADRS reduction,p = 0.02). However, after 1 week of initiating stimulation, three of six patients had a > 50% decrease in MADRS scores relative to baseline (43% mean MADRS reduction,p = 0.005). One patient withdrew from study participation. At 52 weeks, four of remaining five patients have > 70% decrease in MADRS scores relative to baseline (73% mean MADRS reduction,p = 0.007). Evaluation of modulated fiber tracts reveals significant common orbitofrontal connectivity to the target region in all responders. Neuropsychological testing and18F-fluoro-deoxyglucose-positron emission tomography cerebral metabolism evaluations performed at baseline and at 52 weeks showed minimal changes and verified safety. This longitudinal evaluation of MFB DBS demonstrated rapid antidepressant effects, as initially reported by Schlaepfer et al. (2013), and supports the use of DBS for TRD.
Object: Wound complications are uncommon following deep brain stimulation (DBS) surgery. However, certain key technical steps can be performed in each procedure to minimize this still troublesome risk. The authors reviewed the incidence and management of all hardware-related wound dehiscences and infections in a large patient series. Methods: All patients undergoing new DBS hardware implantation surgery between 2002 and 2010 by a single surgeon (R.K.S.) were entered into a database after undergoing verification by cross-referencing manufacturer implantation records. All hardware-related complications such as wound dehiscence, erosions, and/or infections were identified, and wound location, time of incidence, and mechanism were categorized. Charts were reviewed to evaluate the success of conservative treatment versus partial or total hardware removal. Results: Seven hundred twenty-eight patients received 1333 new DBS leads and 1218 new implantable pulse generators (IPGs) in a total of 1356 stereotactic procedures for movement disorders. Seventy-eight percent of patients underwent staged lead and IPG implantations. Sixteen patients presented with atraumatic device-related infection and/or dehiscence within 12 months from original implantation; 9 of these patients (1.24%) required additional surgery after antibiotic failure. All 8 patients presenting with cranial wound complications were treated initially by debridement in an attempt to salvage the leads; debridement followed by intravenous antibiotics was only successful in 2 patients presenting with dehiscence alone. One of 2 lead-only removals was successful in infections originating in the cranium; the only IPG-originating infection was treated by partial hardware removal and intravenous antibiotics. Two of 637 IPG replacements resulted in infections within 12 months after revision, requiring either partial or total hardware removal, while 1 dehiscence in this group was treated by debridement alone. Conclusions: In a large series of new DBS hardware implantations, the incidence of postoperative wound dehiscence and/or infections requiring further surgery was 1.24%. Standard practice for all implantations was a short procedural duration, copious povidone-iodine irrigation, and postoperative antibiotic administration. Partial hardware removal should be initially attempted for infection. Debridement alone is successful in treating dehiscence without infection.