Vocal folds are an essential part of human voice production. The biomechanical properties are a good indicator for pathological changes. In particular, as an oscillation system, changes in the biomechanical properties have an impact on the vibration behavior. Subsequently, those changes could lead to voice-related disturbances. However, no existing examination combines biomechanical properties and spatial imaging. Therefore, we propose an image registration-based approach, using ultrasound in order to gain this information synchronously. We used a quasi-static load to compress the tissue and measured the displacement by image registration. The strain distribution was directly calculated from the displacement field, whereas the elastic properties were estimated by a finite element model. In order to show the feasibility and reliability of the algorithm, we tested it on gelatin phantoms. Further, by examining ex vivo porcine vocal folds, we were able to show the practicability of the approach. We displayed the strain distribution in the tissue and the elastic properties of the vocal folds. The results were superimposed on the corresponding ultrasound images. The findings are promising and show the feasibility of the suggested approach. Possible applications are in improved diagnosis of voice disorders, by measuring the biomechanical properties of the vocal folds with ultrasound. The transducer will be placed on the vocal folds of the anesthetized patient, and the elastic properties will be measured. Further, the understanding of the vocal folds’ biomechanics and the voice forming process could benefit from it.
A measurement setup combined with a numerical simulation by a linear finite element analysis is presented as a method to determine the elastic modulus of both artificial and real tissue as a function of frequency. At the end, the future goal is to develop and validate the method to measure the elastic modulus of in-vivo human vocal folds over the human phonation frequency range. In the present study, a miniaturized acoustic pressure pipette aspiration technique is developed to measure the material characteristics of an isotropic silicone specimen with similar characteristics as human vocal folds. In previous studies, friction and compression force effects of the pipette tip wall on the surface of the sample and the radius of the pipette were not investigated. Moreover, the large scale of the measurement setups made them impossible to use for clinical applications. Therefore, two different pipette sample cross-section boundary conditions and two different pipette radii were used. With the aim of ensuring reliable results, we tested our method with pipettes of two different radii on four silicone samples with different consistencies over a frequency range of 50–500 Hz. The simulation verified the measurement results in which the strong dependency of the elastic modulus on the excitation frequency, radius of the pipette, the pipette tip compression force and friction was revealed. By the simulation results, two different frequency dependent equations were developed for calculating elastic modulus of the silicone mixtures in the two cross-section boundary conditions. It was concluded that using a very small gap in between the pipette tip and the specimen can cancel the impact of the pipette tip force and friction which are the major cause of uncertainty. However, if a connection between the pipette and the surface is unpreventable, the contact force should be restricted to be absolutely zero.