The construction and first proton beam tests of a demonstrator dedicated to the beam ballistic control in hadrontherapy cancer treatments are described. This cost-effective demonstrator, called large area pixelized detector, is a PET-like detector used for in-beam ballistic control. It was built to test the feasibility of monitoring in real time, during irradiation, the ion range in the patient through the measurement of the beam-induced β + activity distribution. Achieving this goal necessitates to overcome several challenges. One of them is the rejection of the beam-induced background. Another one is the definition of fast event selection and reconstruction techniques so that real time monitoring is possible. Strategies employed to tackle these problems are presented and tested with the 65 MeV Medicyc proton beam of the cancer treatment center in Nice, France. In particular, an original fast reconstruction technique is presented. First performances obtained during irradiation of polymethyl methacrylate targets are described.
Proton therapy (PRT) is an irradiation technique that aims at limiting normal tissue damage while maintaining the tumor response. To study its specificities, the ARRONAX cyclotron is currently developing a preclinical structure compatible with biological experiments. A prerequisite is to identify and control uncertainties on the ARRONAX beamline, which can lead to significant biases in the observed biological results and dose–response relationships, as for any facility. This paper summarizes and quantifies the impact of uncertainty on proton range, absorbed dose, and dose homogeneity in a preclinical context of cell or small animal irradiation on the Bragg curve, using Monte Carlo simulations. All possible sources of uncertainty were investigated and discussed independently. Those with a significant impact were identified, and protocols were established to reduce their consequences. Overall, the uncertainties evaluated were similar to those from clinical practice and are considered compatible with the performance of radiobiological experiments, as well as the study of dose–response relationships on this proton beam. Another conclusion of this study is that Monte Carlo simulations can be used to help build preclinical lines in other setups.
This paper describes a small prototype of an in beam PET like detector, named ”Large Acceptance Pixelized Detector” (LAPD), developed to test technical concepts for the ion range control in the context of cancer treatments using proton or ion beams. The mechanical characteristics of this detector together with the read-out electronics are first presented. Then, results of a first experiment, performed on a 65 MeV proton beamline, are reported. Finally, we discuss the ability of Geant4 Monte-Carlo to reproduce the experimental data.