Huong Nguyen Minh

Mrs. Huong Nguyen Minh

Charité - Universitätsmedizin Berlin, Berlin, Germany

Research Focus & Metrics

Academic Career Length

8Years (2019-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
24 October 2019 in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
Verified
Journal Article
Open access
HMHuong Nguyen Minh
Huong Nguyen Minh
JDJuan Du
Juan Du
KRKay Raum
Kay Raum
Most bone loss during the development of osteoporosis occurs in cortical bone at the peripheral skeleton. Decreased cortical thickness (Ct.Th) and the prevalence of large pores at the tibia are associated with reduced bone strength at the hip. Ct.Th and cortical sound velocity, i.e., a surrogate marker for changes of cortical porosity (Ct.Po), are key biomarkers for the identification of patients at high fracture risk. In this study, we have developed a method using a conventional ultrasound array transducer to determine thickness (Ct.Th) and the compressional sound velocity propagating in the radial bone direction (Ct. ν11 ) using a refraction-corrected multifocus imaging approach. The method was validated in-silico on porous bone plate models using a 2-D finite-difference time-domain method and ex vivo on plate-shaped plastic reference materials and on plate-shaped cortical bovine tibia samples. Plane-wave pulse-echo measurements provided reference values to assess precision and accuracy of our method. In-silico results revealed the necessity to account for inclination-dependent transmission losses at the bone surface. Moreover, the dependence of Ct. ν11 on both porosity and pore density was observed. Ct.Th and Ct. ν11 obtained ex vivo showed a high correlation ) with reference values. The ex-vivo accuracy and precision for Ct. ν11 were 29.9 m/s and 0.94%, respectively, and those for Ct.Th were 0.04 mm and 1.09%, respectively. In conclusion, this numerical and experimental study demonstrates an accurate and precise estimation of Ct.Th and Ct. ν11 . The developed multifocus technique may have high clinical potential to improve fracture risk prediction using noninvasive and nonionizing conventional ultrasound technology with image guidance.

9 August 2021 in JBMR Plus
Verified
Journal Article
Open access
PDF file
GAGabriele Armbrecht
Gabriele Armbrecht
HMHuong Nguyen Minh
Huong Nguyen Minh
JMJonas Massmann
Jonas Massmann
KRKay Raum
Kay Raum
Osteoporosis is a disorder of bone remodeling leading to reduced bone mass, structural deterioration, and increased bone fragility. The established diagnosis is based on the measurement of areal bone mineral density by dual‐energy X‐ray absorptiometry (DXA), which poorly captures individual bone loss and structural decay. Enlarged cortical pores in the tibia have been proposed to indicate structural deterioration and reduced bone strength in the hip. Here, we report for the first time the in vivo assessment of the cortical pore‐size distribution together with frequency‐dependent attenuation at the anteromedial tibia midshaft by means of a novel ultrasonic cortical backscatter (CortBS) technology. We hypothesized that the CortBS parameters are associated with the occurrence of fragility fractures in postmenopausal women (n = 55). The discrimination performance was compared with those of DXA and high‐resolution peripheral computed tomography (HR‐pQCT). The results suggest a superior discrimination performance of CortBS (area under the receiver operating characteristic curve [AUC]: 0.69 ≤ AUC ≤ 0.75) compared with DXA (0.54 ≤ AUC ≤ 0.55) and a similar performance compared with HR‐pQCT (0.66 ≤ AUC ≤ 0.73). CortBS is the first quantitative bone imaging modality that can quantify microstructural tissue deteriorations in cortical bone, which occur during normal aging and the development of osteoporosis. © 2021 The Authors. JBMR Plus published by Wiley Periodicals LLC on behalf of American Society for Bone and Mineral Research.

23 May 2022 in Applied Sciences
Verified
Journal Article
Open access
PDF file
HMHuong Nguyen Minh
Huong Nguyen Minh
MMMARIE MULLER
MARIE MULLER
KRKay Raum
Kay Raum
Delay-and-sum (DAS) beamforming of backscattered echoes is used for conventional ultrasound imaging. Although DAS beamforming is well suited for imaging in soft tissues, refraction, scattering, and absorption, porous mineralized tissues cause phase aberrations of reflected echoes and subsequent image degradation. The recently developed refraction corrected multi-focus technique uses subsequent focusing of waves at variable depths, the tracking of travel times of waves reflected from outer and inner cortical bone interfaces, the estimation of the shift needed to focus from one interface to another to determine cortical thickness (Ct.Th), and the speed of sound propagating in a radial bone direction (Ct.ν11). The method was validated previously in silico and ex vivo on plate shaped samples. The aim of this study was to correct phase aberration caused by bone geometry (i.e., curvature and tilt with respect to the transducer array) and intracortical pores for the multi-focus approach. The phase aberration correction methods are based on time delay estimation via bone geometry differences to flat bone plates and via the autocorrelation and cross correlation of the reflected ultrasound waves from the endosteal bone interface. We evaluate the multi-focus approach by incorporating the phase aberration correction methods by numerical simulation and one experiment on a human tibia bone, and analyze the precision and accuracy of measuring Ct.Th and Ct.ν11. Site-matched reference values of the cortical thickness of the human tibia bone were obtained from high-resolution peripheral computed tomography. The phase aberration correction methods resulted in a more precise (coefficient of variation of 5.7%) and accurate (root mean square error of 6.3%) estimation of Ct.Th, and a more precise (9.8%) and accurate (3.4%) Ct.ν11 estimation, than without any phase aberration correction. The developed multi-focus method including phase aberration corrections provides local estimations of both cortical thickness and sound velocity and is proposed as a biomarker of cortical bone quality with high clinical potential for the prevention of osteoporotic fractures.