Cryoactivation is known to occur in whole blood and plasma samples when kept between +4 and −5 °C, leading to falsely high renin concentrations. In 2022 it has been clearly shown that cryoactivation can also occur in samples stored at −20 °C. Based on these new findings, here we discuss how this can influence the clinical diagnosis of patients. First, we show that storage of renin plasma samples can affect the renin measurements and thereby the aldosterone to renin ratio (ARR) calculation, which might explain the high intraindividual variability in ARR also recently demonstrated. Second, we discuss the existing studies on the establishment of renin reference intervals and note the lack of attention given to this recently revealed preanalytical condition. Our literature review of the reference intervals for renin suggest that cryoactivation might have influenced the published data.
1 October 2010 in Journal of Clinical Psychopharmacology
Background: Accurate and precise pipetting is critical for reliable laboratory results, especially when handling small volumes or sensitive analytical techniques. Despite the common use of piston-operated pipettes, operator-dependent variability remains a major source of error. This study aimed to evaluate pipetting performance among laboratory personnel with varying levels of experience. Methods: The study included 108 participants: 26 laboratory school students (first to third year), 10 PhD students, 10 clinical chemists, and 62 laboratory technicians. Technicians were stratified into 3 groups based on the extent of sensitive manual pipetting in their work: minimal (Group A), moderate (Group B), and high (Group C). Pipetting performance was evaluated using gravimetric replicates at 10 µL and 100 µL volumes, and a dilution task involving serial dilutions of 1 M copper sulfate. Accuracy and precision metrics were compared across groups. Results: The median coefficient of variation (CV%) was lower for 100 µL replicates (0.3% to 0.8%) than for 10 µL replicates (1.4% to 5.9%). Laboratory students showed the greatest variability and longest task completion times, although performance improved with advancing school year. Ten percent of participants exceeded ISO-defined accuracy limits for 10 µL pipetting, and 22% for 100 µL. Imprecision limits were surpassed by 36% and 20% of participants for 10 µL and 100 µL, respectively. Group C technicians achieved optimal dilution performance (R2 = 1.00), while students showed the highest deviation. Conclusions: Pipetting performance was significantly influenced by experience level and task sensitivity. Even experienced personnel occasionally failed to meet ISO standards, underscoring the need for ongoing training and performance assessment.
Exposure of plasma samples to low temperatures (between -5 and + 4 °C) leads to conversion of prorenin to its open state, which is then converted to renin by plasma proteases. This process, known as cryoactivation, results in falsely elevated renin concentrations. This study examined the effects of storage temperature, protease inhibitors, and 2-hour incubation at 37 °C on renin cryoactivation. Blood samples (n = 24) were collected in four types of tubes: K2EDTA, aprotinin, trypsin inhibitor, and protease inhibitor cocktail. Renin concentrations (IDS-i10) were measured at baseline and after storage at + 4 °C (one week), -20 °C (two weeks), and -80 °C (two weeks). Samples stored at -20 °C were later incubated at 37 °C for two hours and reanalyzed. Aldosterone concentrations were measured (LC-MS/MS) and aldosterone to renin ratios (ARR) were calculated. Renin concentrations remained stable at -80 °C but increased significantly at + 4 °C (18-50 %) and -20 °C (61-500 %). ARR changes were minimal at -80 °C (-8 pmol/mU) but largest at -20 °C (-263 pmol/mU). Renin concentrations returned to baseline after a 2-hour incubation at 37 °C only in trypsin inhibitor and protease inhibitor cocktail tubes. Our study shows more cryoactivation of renin in plasma aliquots stored at -20 °C compared to storage at + 4 °C. Cryoactivation of prorenin in trypsin inhibitor and protease inhibitor cocktail tubes was reversed by a 2-hours incubation at 37 °C, which may be an alternative procedure for laboratories that are not able to store renin samples in -80 °C freezers.
26 November 2023 in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
This case report describes the positive interference of the commonly used skin protective barrier cream used together with urine collection bags on the benzethonium chloride method for urine protein measurements in a 6-month-old female baby, leading to falsely elevated results. The interference was identified by both artificially mixing urine samples with this cream and comparing the results obtained using the benzethonium chloride method with those obtained using the pyrogallol red method.