This protocol describes a method for the routine operation of the Agilent Triple Quadrupole Inductively Coupled Plasma Mass Spectrometer 8900 (ICP-QQQ) for the elemental analysis of digested cells, spent media, and reagents. This protocol will outline the formulation of needed solutions, startup checklist, MassHunter software operation, performance checks, tuning, calibration, sample introduction, data acquisition, data analysis, and shut down. The parameters used in this protocol enable the efficient and accurate quantitation of Na, Mg, P, S, K, Ca, Fe, Mn, Co, Ni, Cu, Zn, Se, Mo, Ag, and Cd in algal cells, the growth media used for culture, and spent media. Cells digested in 70% nitric acid or the media are separately diluted to a final matrix of 2% nitric acid. Samples are mixed with on-line addition of internal standards Sc, Y, and Re. Three gases are used sequentially in the collision reaction cell to minimize interferences. Helium is used in collision mode for the removal of polyatomic ions through kinetic energy discrimination. Hydrogen and oxygen are used in reaction mode for the removal of mass shifted contaminants and the selection of mass shifted ions, respectively. The linear dynamic range of the instrument using this set up spans 4 orders of magnitude with a detection limit in the parts per trillion range for trace elements such as Cu, Zn, Mn, Se, Mo, Ag, and Cd.