22 March 2019 in IEEE Transactions on Device and Materials Reliability
Verified
Journal Article
Closed access
Jakob JantingJakob Janting
Jakob Janting
JTJulie Grundtvig Theander
Julie Grundtvig Theander
HEHenrik Egesborg
Henrik Egesborg
An empirical rule, the 10 ∘C rule, states that chemical reaction rates are doubled for every 10 ∘C temperature increase. This is often used in thermally accelerated medical device polymer aging studies. Here, theoretical evidence and limitations for the rule are analyzed. Thus, a new more accurate rule based on averaging Arrhenius chemical reaction rate ratios over typical activation energies 0.1 eV -0.9 eV in the normal medical device accelerated test temperature interval 25 ∘C -70 ∘C is proposed. Comparison with the 10 ∘C rule shows that the 10 ∘C rule provides similar estimates, but only at the reference temperature 25 ∘C. Fitting the reaction rate ratio based on the Arrhenius equation using the reference temperature 25 ∘C to the 10 ∘C rule data reveals that best agreement is achieved with a thermal aging activation energy of 0.67 eV.