In this paper, we present a single-shot dual-mode imaging system that uses optical ultrasound (US) as an ultrasonic pulser; thus without transmission circuit. The ultrasonic pulse-echo system comprises an optical US pulser generated by carbon nanotubes (CNTs), which generate a high-power photoacoustic (PA) signal, and a capacitive micromachined ultrasonic transducer (CMUT) receiver. By fabricating a thin CNT-polydimethylsiloxane (PDMS) composite capable of semi-absorption of the laser, a single-shot imaging system was developed. By transmitting semi-penetration light to the object, US and PA imaging were performed in a single shot. A CNT thickness of <1 μm produced a maximum pressure of 154 kPa, and US was received by the CMUT with a 2-MHz center frequency in the PDMS. Additionally, a low-profile and near-depth imaging system was constructed with an intermediate layer of 6-mm PDMS for the dry contact method. We performed a single-shot dual-mode imaging experiment on point and line phantoms, as well as the particle spread in soft tissue. Thus, we examined the feasibility of the near-depth and single-shot dual-mode (US and PA) imaging system capable of dry contact.